9. Troubleshooting
PROGRAMMABLE AC POWER SOURCE
232
Table 9-2 Error Message List (Continued)
ID Message Protection
type
Cause or Necessary action
32
Calibration Data Error ME Message
display only
The adjustment data is abnormal. Cycle the
power. If it still occurs, contact the seller
(NF Corporation or our agent) from whom
you purchased the product.
33 Line
Overvoltage
System
Lock
The power input voltage is excessive. Check
if the power input is appropriate.
34 Line
Undervoltage
System
Lock
The power input voltage is insufficient.
Check if the power input is appropriate.
36 Communication
Failure
PU
System
Lock
There is a communication error between the
firmware and the power unit. Cycle the
power. If it still occurs, contact the seller
(NF Corporation or our agent) from whom
you purchased the product.
37 Communication
Failure
ME
System
Lock
There is a communication error between the
firmware and the output measurement
section. Cycle the power. If it still occurs,
contact the seller (NF Corporation or our
agent) from whom you purchased the
product.
43
Mismatched Pw Unit
Versions
System
Lock
All power unit versions do not match.
44
Unsupported Pw Unit
Version
System
Lock
A power unit of unsupported version is
connected.
46 Inhibited
Cabinet
Connect System
Lock
A multi-phase model is connected with
system cable. Use the multi-phase model as
a single unit.
47
No Available Power Unit
System
Lock
There is no power unit that can be started.
48
Output Overvoltage
Output Off
The output voltage is excessive. This may
occur at an abrupt change in the output
current due to the inductive load.
Also when the remote sensing function set
to FB, check the feedback cable connects
correctly. (See
8.2.2
)
50
Output Overcurrent [RMS] Output Off
The output current RMS value is excessive
on a module in the power unit. This may
occur when the output terminal is
short-circuited.
Also when the remote sensing function set
to FB, check the feedback cable connects
correctly. (See
8.2.2
)
51
Output Overcurrent [Peak] Output Off
The output current peak value is excessive.
52
Power Unit DCPS Error
Output Off
The DC power supply in the power unit is
abnormal, or see the notes in "
9.1.3
,
Operation procedure."
Содержание DP060LM
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Страница 4: ......
Страница 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Страница 24: ......
Страница 215: ...DP060LM DP120LM DP180LM 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202...
Страница 224: ......
Страница 225: ...DP060LM DP120LM DP180LM 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 216...
Страница 239: ...DP060LM DP120LM DP180LM 221 8 Peripherals 8 1 Peripherals Line up 222 8 2 Using in Combination with DIP or RIN 223...
Страница 244: ......
Страница 264: ......
Страница 298: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 280 11 35 Outline Dimensional Drawing Figure 11 2 DP060LM Type2L cabinet...
Страница 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...
Страница 307: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...