9.1 Error Messages and Error Handling
DP060LM/DP120LM/DP180LM
229
9.1.3 If
an
Error
Occurs Repeatedly
If an error message repeatedly appears even though you removed the cause of the error, the error may
be resolved by restarting the power unit.
If the error still exists, this product or the power unit in the product may be faulty. Repair is needed
in this case. Contact the seller (NF Corporation or our agent) from whom you purchased the product.
Even though one of the power units is faulty, the product can be used by disconnecting the faulty
power unit and using the rest of them.
The following shows how to restart and disconnect the power unit in the case of an error.
--------
Notes
------------------------------------------------------------------------------------------------------------
The following procedure is available only when the [PU-ON] soft-key is displayed on the error
message window. Stop using this product in the case of a repeating error without the [PU-ON]
soft-key.
-----------------------------------------------------------------------------------------------------------------------------
Operation procedure
1.
In the error message window, press the [PU-ON] soft-key. The power unit energization setting
window opens.
2.
The power unit with an error is marked with "E" like
. Check on the power unit that has an
error. First, leaving the energization setting as is, select OK. All the power units are restarted.
3.
When the error message window is displayed again, press the ENTER key.
4.
If the error still exists, press the [PU-ON] soft-key again to open the power unit energization
setting window.
5.
Check on the power unit that has an error. If the same power unit still has an error, it may be
faulty. Move the cursor to this power unit, then press the ENTER key. The check mark on this
power unit is removed, showing
. If a different power unit has an error, leave the check mark
as is. Select [OK].
6.
The power unit energization setting is updated, and only the power units with the check mark
are restarted. When the error message window is displayed again, press the ENTER key.
7.
Repeat the step 4 and later ones until errors are resolved. After the error is resolved, the product
can be continuously used using the power units with the check mark.
Содержание DP060LM
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Страница 4: ......
Страница 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Страница 24: ......
Страница 215: ...DP060LM DP120LM DP180LM 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202...
Страница 224: ......
Страница 225: ...DP060LM DP120LM DP180LM 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 216...
Страница 239: ...DP060LM DP120LM DP180LM 221 8 Peripherals 8 1 Peripherals Line up 222 8 2 Using in Combination with DIP or RIN 223...
Страница 244: ......
Страница 264: ......
Страница 298: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 280 11 35 Outline Dimensional Drawing Figure 11 2 DP060LM Type2L cabinet...
Страница 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...
Страница 307: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...