4.
Advanced Operation
PROGRAMMABLE AC POWER SOURCE
164
4.13 Using
Autocal
Function
The Autocal (Automatic Calibration) function calculates the ratio (calibration factor) of the output
voltage measured value (effective value) versus the output voltage setting value when this function is
turned on, and multiplies the gain of the output amplifier by the calculated value, intending to match
the output voltage with the setting value. The calculated calibration factor is used as a fixed value
until the Autocal function is turned off. Therefore, even when the Autocal function is on, the output
voltage fluctuation may occur if the load is fluctuated. When the Autocal function is set to on, the
icon
is displayed.
As the detected part of the output voltage calibrated by the Autocal function, either the sensing input
terminal (the Remote sensing function is on) or the output terminal (the Remote sensing function is
off) can be selected. Combining the Remote sensing function with the Autocal function, you can
compensate the voltage drop due to wiring to the load.
The Autocal function can be set to on in the AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA
modes. However, for the AC mode, it can be set to on only when the waveform is a sine wave. When
the Autocal function is on, the waveform cannot be changed to other than a sine wave. When the
mode is changed to ACDC mode, when the signal source is changed to EXT or ADD, or when the
Sequence or Simulation function is selected, the Autocal function is forcibly set to off. When the
AGC function is on or Remote sensing function is FB, the Autocal function cannot be used.
However, in case the stand-by state of sequence mode (step0) effective only for AC-INT, ACDC-INT
or DC-INT and when the waveform is sine wave or DC. Autocal cannot be activated unless either AC
voltage or DC voltage setting is not 0 V for ACDC-INT. Also, Autocal is turned off once when you
change ACDC mode from AC mode of DC mode even in stand-by state. Turn ON it again when you
need.
For the specification of the Autocal function, see
11.18
.
Содержание DP060LM
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Страница 4: ......
Страница 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Страница 24: ......
Страница 215: ...DP060LM DP120LM DP180LM 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202...
Страница 224: ......
Страница 225: ...DP060LM DP120LM DP180LM 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 216...
Страница 239: ...DP060LM DP120LM DP180LM 221 8 Peripherals 8 1 Peripherals Line up 222 8 2 Using in Combination with DIP or RIN 223...
Страница 244: ......
Страница 264: ......
Страница 298: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 280 11 35 Outline Dimensional Drawing Figure 11 2 DP060LM Type2L cabinet...
Страница 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...
Страница 307: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...