4.11 Using Remote Sensing Function
DP060LM/DP120LM/DP180LM
159
When setting Remote sensing to FB, the load cable connected to the output terminal should be
as short as possible (3 m or less) with twisted or parallel wire.
Cable connection for remote sensing
Before connecting the cable, turn off this product including its peripherals.
Pay attention to the withstand voltage of the cable to be connected.
Do not connect anything to the terminals that are not used.
Do not input a signal having a different electric potential from the output terminal (e.g. the
signal at the secondary side of transformer) into the sensing input terminal.
For the polyphase output, cables should be connected to the sensing input terminals N1, N2, and
N3 for each phase. These terminals do not have the same electric potential within this product.
Connect the sensing input terminal to the output voltage detecting end (e.g. load end) using a cable.
The sensing input terminal is common to each output phase mode (single-phase two-wire,
single-phase three-wire, and three-phase four-wire). The terminals to be used for each output phase
mode are shown in Figure 4-24. Do not connect anything to the terminals that are not used.
単相
2
線出力で使用
単相
3
線出力で使用
三相
4
線出力で使用
Used for the single-phase two-wire output
Used for the single-phase three-wire output
Used for the three-phase four-wire output
Figure 4-24 Sensing input terminals
CAUTION
WARNING
CAUTION
Содержание DP060LM
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Страница 4: ......
Страница 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Страница 24: ......
Страница 215: ...DP060LM DP120LM DP180LM 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202...
Страница 224: ......
Страница 225: ...DP060LM DP120LM DP180LM 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 216...
Страница 239: ...DP060LM DP120LM DP180LM 221 8 Peripherals 8 1 Peripherals Line up 222 8 2 Using in Combination with DIP or RIN 223...
Страница 244: ......
Страница 264: ......
Страница 298: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 280 11 35 Outline Dimensional Drawing Figure 11 2 DP060LM Type2L cabinet...
Страница 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...
Страница 307: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...