4.
Advanced Operation
PROGRAMMABLE AC POWER SOURCE
104
Step
The output by the Simulation function consists of six types of steps: Initial, Normal 1, Trans 1,
Abnormal, Trans 2, and Normal 2. Before the simulation starts, this product stands by at the Initial
Step. During the simulation, the step changes in the order of Initial, Normal 1, Trans 1, Abnormal,
Trans 2, and Normal 2. After the simulation ends, this product returns to the Initial Step and stands
by.
Step parameter
Each parameter of Step Time, AC voltage, frequency, start phase, stop phase, trigger output, and
synchronization output can be set for each step. However, some parameters cannot be set in some
steps.
Initial Step
Initial Step is assigned to the stand-by state before the simulation starts. This product also moves to
the Initial Step and becomes stand-by state after the simulation ends, except when the repeat count is
specified to 0 (= infinite count).
Transition Step (Trans 1 and Trans 2)
The Step Times of Trans 1 and Trans 2 can be set to zero for simulating the abrupt voltage and phase
changes. When the Step Time is set to the value other than zero, the set voltage and frequency are
swept linearly at the previous and next steps. Note that the AC voltage, frequency, start phase, and
Stop Phase cannot be set in the Transition Steps.
Other issues
The Exit operation is available during the simulation. Performing the Exit operation moves to
the Initial Step. The output remains on.
You can set the repeat on/off and repeat count for the simulation. When the repeat is set to on
and the repeat count to zero (0), the testing repeats infinitely.
--------
Notes
------------------------------------------------------------------------------------------------------------
The Simulation function does not support the standard tests defined by organizations such as
IEC. Use this function for a preliminary testing.
-----------------------------------------------------------------------------------------------------------------------------
Содержание DP060LM
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Страница 4: ......
Страница 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Страница 24: ......
Страница 215: ...DP060LM DP120LM DP180LM 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202...
Страница 224: ......
Страница 225: ...DP060LM DP120LM DP180LM 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 216...
Страница 239: ...DP060LM DP120LM DP180LM 221 8 Peripherals 8 1 Peripherals Line up 222 8 2 Using in Combination with DIP or RIN 223...
Страница 244: ......
Страница 264: ......
Страница 298: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 280 11 35 Outline Dimensional Drawing Figure 11 2 DP060LM Type2L cabinet...
Страница 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...
Страница 307: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...