ADVANCED MEASUREMENT
75
Relative Value Measurement
Applicable to
Background
Relative measurement stores a value, typically the data at the moment,
as the reference. The following measurement is shown as the delta
between the references. The reference value will be cleared upon exit.
REL, basically, is to subtract a certain value in the following
measurement. The value is fixed and remains its effect even user exits
and returns back to this function again.
One of the most seen purposes of REL is to eliminate impedance of
test lead from measurement. Before operating impedance
measurement, short circuit the test lead followed by pressing the
[REL] button. For other measurements, press the [REL] button after
putting test lead in a null circuit.
Alternatively, user can modify the value by pressing the [REL#]
button followed by using the knob or number keys to enter a
specified value. Press the [REL] button again to disable null
operation.
Activate
Relative
measurement
Press the REL key. The measurement
reading at the moment becomes the
reference value.
Relative
measurement
display appears
Indicator Relative
Measurement
Relative
Reference Value
REL Delta Value
REL
Indicates Relative value measurement
REL: +000.5711mV
Shows the stored reference value
+000.2653
Shows the delta between the current
measurement data and the reference value
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...