PPA2500 
PPA2600 

 
 

KinetiQ 

 

Communications 

manual 

 
 

 

 

Firmware v1.100 

 

10

th

 April 2017

 

Содержание KinetiQ

Страница 1: ...PPA2500 PPA2600 KinetiQ Communications manual Firmware v1 100 10th April 2017...

Страница 2: ...ent is powered from a properly grounded supply The inputs are rated at 1kV rms or dc cat II 600V rms or dc cat III Do not exceed the rated input Keep the ventilation holes on the underneath and rear f...

Страница 3: ...each command is given in the appendix Although most of the commands apply to all instruments in the range there are some commands that are specific to one instrument or another The information in this...

Страница 4: ...rmat 1 6 2 Communication commands 2 1 CLS 2 1 ESE 2 2 ESR 2 3 IDN 2 4 OPC 2 5 RST 2 6 SRE 2 7 SRE 2 8 STB 2 9 TRG 2 10 TST 2 11 WAI 2 12 ABORT 2 13 ALARM 2 14 ALARM 2 15 ALARM1 2 16 ALARM2 2 17 ALARME...

Страница 5: ...2 45 HOLD 2 46 INPUT 2 47 INTEGR 2 48 INTEGR 2 49 KEYBOA 2 50 LCR 2 51 LCR 2 52 LOWFRE 2 53 MODE 2 54 MSLAVE 2 55 MULTIL 2 56 MULTIL 2 57 NOOVER 2 58 PFCONV 2 59 PHASEM 2 60 PHASEM 2 61 PHCONV 2 62 P...

Страница 6: ...TATUS 2 82 STOP 2 83 SUSPEN 2 84 TEMPER 2 85 TEMPER 2 86 TORQSP 2 87 TORQSP 2 88 USER 2 89 VARCON 2 90 VERSIO 2 91 VRMS 2 92 VRMS 2 93 WIRING 2 94 ZERO 2 95 ZOOM 2 96 ZOOM 2 97 Appendix A command summ...

Страница 7: ...EEE488 interface is set to remote the other ports are ignored The commands are not case sensitive and white space characters are ignored e g tabs and spaces Replies from the instrument are always uppe...

Страница 8: ...LAN see section 5 2 The IEEE address defaults to 23 and can be changed via the COMMS menu The keyboard is disabled when the instrument is set to remote using the IEEE Press HOME to return to local ope...

Страница 9: ...en the instrument has an error bit 4 EXE execution error set when the command cannot be executed bit 5 CME command interpretation error set when a command has not been recognised bit 7 PON power on ev...

Страница 10: ...as enabled by DAVER bit 3 ALA alarm active set when an alarm is active and enabled by ALARMER bit 4 MAV message available set when a message reply is waiting to be read bit 5 ESB standard event summar...

Страница 11: ...ll only transmit when CTS pin 8 is asserted and can only receive if DCD pin 1 is asserted The instrument constantly asserts 12V DTR pin 4 so this pin can be connected to any unwanted modem control inp...

Страница 12: ...ata is sent as a 7 bit signed exponent a mantissa sign and a 20 bit mantissa byte data 1 7 bit signed exponent 63 to 64 2 bit 6 mantissa sign bit 5 0 mantissa bit 19 14 3 mantissa bit 13 7 4 mantissa...

Страница 13: ...25 26xx communications manual 2 1 2 Communication commands CLS CLS Function Clear status Description Clears the standard event status register Format CLS Arguments none Reply none Example CLS ESR 0 No...

Страница 14: ...event status enable register Reply can be read by ESE Example ESE 60 Notes The following bits in the standard event status enable register have been implemented bit 0 OPC operation complete bit 2 QYE...

Страница 15: ...dard event status register Example ESR 33 Notes The following bits in the standard event status register have been implemented bit 0 OPC operation complete bit 2 QYE unterminated query error bit 3 DDE...

Страница 16: ...nction Identify query Description Returns a standard format identification string Format IDN Arguments none Reply An ASCII string in the IEEE488 2 format manufacturer model serial no version Example I...

Страница 17: ...for operation complete Description Returns 1 if previous operation is completed 0 if not Format OPC Arguments none Reply 0 or 1 Example START OPC 0 OPC 0 OPC 1 Notes OPC can be used to indicate when...

Страница 18: ...to the default state and clears the standard event status register Format RST Arguments none Reply none Example RST Notes The RST command loads the default configuration This is the same as loading th...

Страница 19: ...t enable register Description Enable which bits of the status byte register initiate a service request Format SRE value Arguments decimal equivalent of bits in status byte register Reply can be read b...

Страница 20: ...Read service request enable register Description Read back the present setting of the service request enable register Format SRE Arguments Reply decimal equivalent of bits in status byte register that...

Страница 21: ...the serial poll status byte Format STB Arguments none Reply decimal value of the serial poll status byte Example STB 1 Notes The following bits in the serial poll status register have been implemente...

Страница 22: ...5 26xx communications manual 2 10 TRG TRG Function Trigger Description Initiates a new measurement resets the range and smoothing Format TRG Arguments none Reply none Example MODE VRMS TRG VRMS SURG N...

Страница 23: ...11 TST TST Function Self test query Description Returns the results of self test Format TST Arguments none Reply single integer bit 0 set if uncalibrated bit 1 set if error with analogue zero 15 majo...

Страница 24: ...nications manual 2 12 WAI WAI Function Wait for operation complete Description Suspends communication until the previous operation has completed Format WAI Arguments none Reply none Example TRG WAI PO...

Страница 25: ...x communications manual 2 13 ABORT ABORT Function Abort datalog Description Abort datalog data acquisition Format ABORT Arguments none Reply none Example DATALOG RAM 0 02 START wait for data values AB...

Страница 26: ...14 ALARM ALARM Function Set common controls for alarm1 and alarm2 Description Set the alarm latch and sounder control Format ALARM latch sounder Arguments latch ON OFF sounder ENABLED DISABLED Reply n...

Страница 27: ...n Reads the status of the measurements and 2 alarms Format ALARM Arguments none Reply single integer bit 0 data available bit 1 data error bit 2 alarm 1 bit 3 alarm 2 Example ALARM 1 Notes An alarm is...

Страница 28: ...larm1 Description Set alarm1 type and thresholds Format ALARM1 type data high low Arguments type DISABLED HIGH LOW INSIDE OUTSIDE LINEAR data 1 4 high high threshold low low threshold Reply none Examp...

Страница 29: ...larm2 type and thresholds Format ALARM2 type data high low Arguments type DISABLED HIGH LOW INSIDE OUTSIDE data 1 4 for zoom data high high threshold low low threshold Reply None Example ALARM2 LOW 3...

Страница 30: ...if any set the alarm active bits in the status byte Format ALARME value Arguments decimal equivalent of alarm bits bit2 set bit 3 of status byte when alarm 1 is active bit3 set bit 3 of status byte w...

Страница 31: ...tion Read alarm status enable register Description Read back present bits in the alarm status enable register which controls the alarm active bit in the status byte Format ALARME Arguments none Reply...

Страница 32: ...or optimum measurement Format APPLIC type setting Arguments type NORMAL PWM PWM Motor Drive BALLAST Lighting ballast INRUSH Inrush Current POWERT Transformer mode STANDB Standby power CALIBR Calibrati...

Страница 33: ...th can be restricted to low to 40kHz For measurements of dc in the presence of large ac signal the bandwidth can be further restricted to dc only to 10Hz Format BANDWI phase type Arguments phase PHASE...

Страница 34: ...tion Returns a numerical value for the bandwidth setting Format BANDWI phase Arguments phase PHASE1 PHASE2 PHASE3 Reply 0 WIDE 1 LOW 2 DCONLY Example BANDWI PHASE3 LOW BANDWI PHASE3 1 Notes If indepen...

Страница 35: ...PPA25 26xx communications manual 2 23 BEEP BEEP Function Sound the buzzer Description Makes a beep from the instrument Format BEEP Arguments none Reply none Example BEEP Notes...

Страница 36: ...A25 26xx communications manual 2 24 BLANKI BLANKI Function Select blanking Description Enable or disable low value blanking Format BLANKI value Arguments value ON OFF Reply none Example BLANKI OFF Not...

Страница 37: ...ring with the new date The original string is not overwritten but the alternative string is displayed instead Format CALVER string Arguments string is any sequence of printable alpha numeric character...

Страница 38: ...ead on the front panel press SYS and LEFT This shows the date of calibration Users who subsequently verify the accuracy in their own calibration facilities can enter an alternative string with the new...

Страница 39: ...which there may not be a direct command Format CONFIG index data Arguments index is the number of the parameter data is the data for that parameter Reply none Example CONFIG 6 1 set phase convention...

Страница 40: ...sent value of a single parameter Format or CONFIG index CONFIG index Arguments index is the parameter number Reply Value of parameter real or integer as appropriate Example CONFIG 6 read phase convent...

Страница 41: ...he input coupling for a given input channel Format COUPLI phase coupling Arguments phase PHASE1 PHASE2 PHASE3 coupling AC DC ACONLY DCONLY Reply none Example COUPLI PHASE2 AC DC Notes In multi phase a...

Страница 42: ...numerical value for the coupling setting Format COUPLI phase coupling Arguments phase PHASE1 PHASE2 PHASE3 Reply 0 AC DC 1 ACONLY 2 DCONLY Example COUPLI PHASE2 AC DC COUPLI PHASE2 0 Notes In multi ph...

Страница 43: ...al speed Arguments function DISABLE RAM NONVOL RECALL DELETE interval datalog interval in seconds speed HIGH Reply none Example DATALOG NONVOL 10 DATALOG RAM 0 HIGH Notes set interval to 0 to record e...

Страница 44: ...records number of records to return 0 return all new records since last read Reply 3 to 6 data values depending on settings index 1 n elapsed time in hours data1 data2 if stored data3 if stored data4...

Страница 45: ...uments none Reply Decimal equivalent of data available bits bit0 new data available bit1 data available bit2 harmonic series data available bit3 efficiency data available bit6 integration data availab...

Страница 46: ...set the data available bits in the status byte Format DAVER value Arguments decimal equivalent of data available bits bit0 set bit 0 of status byte when new data available bit1 set bit 0 of status by...

Страница 47: ...available enable register Description Read back present setting of the data available enable register which controls the status bits that set the data available bits in the status byte Format DAVER A...

Страница 48: ...on the display so that the zoom data can be used for alarms Format DISPLAY page Arguments page PHASE1 PHASE2 PHASE3 SUM NEUTRAL TOTAL FUNDAMENTAL VOLTAGE CURRENT Reply None Example DISPLAY FUNDAMENTAL...

Страница 49: ...anual 2 37 DISPLAY DISPLAY Function Read the displayed data Description Returns all the values presently on the screen Format DISPLAY Arguments none Reply Multiple floating point values separated by c...

Страница 50: ...Description Selects the data to be used for the efficiency calculation Format EFFICI formula Arguments formula 0 disabled 1 phase 1 phase 2 2 phase 2 phase 1 3 slave master 4 master slave 5 mechanica...

Страница 51: ...EFFICI EFFICI Function Read efficiency result Description Reads back the total and fundamental efficiency results Format EFFICI Arguments none Reply 2 data values separated by commas total fundamenta...

Страница 52: ...mode Description Disables the screen drawing for high speed operation Format FAST value Arguments value ON OFF Reply none Example FAST ON Notes FAST mode does not suppress the data acquisition which c...

Страница 53: ...alue Format FQLOCK value Arguments value ON OFF Normal Constant Dynamic Reply none Example FQLOCK ON FQLOCK Dynamic FQLOCK Normal Notes To fix the analysis to a specified frequency first lock the freq...

Страница 54: ...g the frequency Format FQREF phase FQREF channel FQREF phase channel Arguments channel voltage current speed phase PHASE1 PHASE2 PHASE3 Reply none Example FQREF CURRENT Notes Measured phase is always...

Страница 55: ...y Description Sets the analysis frequency in Hz for frequency lock mode Format FREQUE frequency Arguments frequency in Hz Reply none Example FQLOCK ON FREQUE 5e4 set frequency to 50kHz Notes Lock the...

Страница 56: ...ce formula THD THDS harmonic series THD TIF Telephone Influence Factor THF Telephone Harmonic Factor TDD Total Demand Distortion TRD Total Rated Distortion HPHASE Series harmonic phase harmonic indivi...

Страница 57: ...d by DAV Format or or or HARMON HARMON phase HARMON SERIES HARMON phase SERIES Arguments phase PHASE1 PHASE2 PHASE3 NEUTRAL PHASES Reply or or 11 data values separated by commas freq mag1 mag2 hmag1 h...

Страница 58: ...old Description Turns data hold on or off Useful for reading data from different phases without it being changed between reads Format HOLD state Arguments state ON OFF Reply none Example HOLD ON POWER...

Страница 59: ...Set input mode Description Selects the input type of the instrument Format INPUT channel type Arguments channel CH1 CH2 type INTERN EXTATT EXTSHU Reply none Example INPUT CH1 EXTSHU Notes CH1 applies...

Страница 60: ...Watts and current use signed or unsigned values and whether accumulated or averaged values are computed Also sets up run time for integration over a specific interval Format INTEGR type display INTEGR...

Страница 61: ...the selected phase Format INTEGR phase Arguments phase PHASE1 PHASE2 PHASE3 PHASES SUM Reply 13 values separated by commas time Wh WH f VAh VAh f VArh Varh f pf pf f V V f Ah Ah f Example START wait f...

Страница 62: ...anel keyboard Description The front panel keyboard can be disabled to prevent accidental operation Format KEYBOARD value Arguments value ENABLE DISABLE Reply none Example KEYBOARD DISABLE Notes The ke...

Страница 63: ...munications manual 2 51 LCR LCR Function Set LCR meter mode Description Set LCR mode and conditions Format LCR parameter Arguments parameter AUTO CAPACITANCE INDUCTANCE IMPEDANCE Reply none Example LC...

Страница 64: ...ta if necessary Clears new data available bit read by DAV Format LCR phase Arguments phase PHASE1 PHASE2 PHASE3 PHASES Reply 11 data values separated by commas freq Vmag Amag impedance phase R C L tan...

Страница 65: ...ow frequency option for extending the measurement window Format LOWFRE value Arguments value ON OFF Reply none Example LOWFRE ON Notes LOWFRE is mainly used for measuring low frequencies 5 Hz However...

Страница 66: ...mode Description Sets the basic operating mode of the instrument Format MODE type Arguments type POWER power meter INTEGR integrator HARMON harmonic analyser RMS rms voltmeter LCR LCR meter SCOPE osci...

Страница 67: ...E Function Set master slave mode Description Enables the instrument to synchronise with a second instrument to simultaneously measure up to 6 phases Format MSLAVE type Arguments type DISABLE MASTER SL...

Страница 68: ...unction Arguments index 0 clear all 1 64 select data 1 64 phase 1 3 phase 1 3 4 sum 5 neutral function 1 99 see appendix C Reply none Example MULTIL 0 MULTIL 1 1 2 phase 1 Watts MULTIL 2 2 2 phase 2 w...

Страница 69: ...lected by the MULTIL command Multiple reply strings each containing the same number of data values maximum of 64 as selected by the MULTIL command Example MULTIL 0 MULTIL 1 1 2 phase 1 Watts MULTIL 2...

Страница 70: ...om blanking out results in manual ranging Format NOOVER value Arguments value ON OFF Reply none Example NOOVER ON Notes This can be useful when testing devices in a noisy environment The range can be...

Страница 71: ...iven a sign depending convention either negative if lagging current negative if leading current Format PFCONV type Arguments type NEGLAG NEGLEA Reply none Example PFCONV NEGLAG Notes An inductive load...

Страница 72: ...ns manual 2 60 PHASEM PHASEM Function Set phase meter mode Description Select phase meter mode and reference Format PHASE reference Arguments reference CH1 ratio ch2 ch1 CH2 ratio ch1 ch2 Reply none E...

Страница 73: ...data available bit read by DAV Format PHASEM PHASEM phase Arguments phase PHASE1 PHASE2 PHASE3 PHASES Reply 5 data values separated by commas freq mag1 mag2 dB phase Example PHASEM CH1 PHASEM PHASE1...

Страница 74: ...Function Set phase convention Description Set phase convention Format PHCONV convention Arguments convention 180 180 to 180 360 0 to 360 360 0 to 360 Reply none Example PHCONV 360 Notes 0 to 360 degr...

Страница 75: ...ns manual 2 63 POWER POWER Function Set up power analyser mode Description Configure power analyser with sum current display type Format POWER sum type Arguments sum type TOTAL AVERAGE Reply none Exam...

Страница 76: ...Clears new data available status bit Format POWER phase results Arguments phase PHASE1 PHASE2 PHASE3 PHASES SUM NEUTRAL current only results WATTS VOLTAGE CURRENT VECTORS RMS WVA PH PH Reply WATTS fre...

Страница 77: ...cations manual 2 65 PH PH freq rms1 mag1 lag1 rms2 Example POWER VECTORS data returned Notes POWER without specifying the phase returns the appropriate single phase data PHASES returns the data for al...

Страница 78: ...volatile program stores Description Recall store or delete non volatile program store Format PROGRA function number Arguments function RECALL STORE DELETE number 0 100 Reply none Example PROGRA RECALL...

Страница 79: ...ions manual 2 67 PROGRA PROGRA Function Identify current program Description Reads the name of the last program to be loaded or recalled Format PROGRA Arguments none Reply text string Example PROGRA f...

Страница 80: ...ge control for a given input channel Format RANGE channel ranging range Arguments channel CH1 CH2 ranging AUTO UPAUTO MANUAL range range number 1 8 Reply none Example RANGE CH2 MANUAL 4 Notes CH1 sets...

Страница 81: ...RESOLU format Arguments format NORMAL 5 digit mantissa HIGH 6 digit mantissa BINARY compressed format Reply none Example RESOLU HIGH Notes Data format for NORMAL is 1 2345E 00 Data format for HIGH is...

Страница 82: ...tores Description Recall store or delete non volatile results Format RESULT function number Arguments function RECALL STORE DELETE number 1 20 Reply none Example RESULT RECALL 13 Notes There are 3 typ...

Страница 83: ...res Description Reads a directory of the 20 non volatile result locations Format RESULT Arguments none Reply 20 integers separated by commas Example RESULT 0 0 1 3 1 1 0 2 1 1 0 0 0 0 0 0 0 0 0 0 Note...

Страница 84: ...ommunications manual 2 72 REZERO REZERO Function Rezero front end Description Request the DSP to re compensate for dc offset and compute a new autozero Format REZERO Arguments none Reply none Example...

Страница 85: ...r Description Set a multiplying scale factor for a given input channel Format SCALE channel factor Arguments channel CH1 CH2 factor multiplying scale factor Reply none Example SCALE CH2 10 Notes CH1 s...

Страница 86: ...w oscilloscope data Format SCOPE channel SCOPE phase channel Arguments phase PHASE1 PHASE2 PHASE3 NEUTRA channel VOLTAGE CURRENT Reply 252 signed integers range trigger 250 x data Example HOLD ON SCOP...

Страница 87: ...at Format SCREEN Arguments none Reply Multiple data bit values Example SCREEN data returned Notes SCREEN response ASCII coded Hex 2 characters for each byte 240 lines of 40 bytes each line represents...

Страница 88: ...e data back to the instrument As it sends all settings in a compressed format it is quicker than setting individual parameters Format SETUP index data Arguments index 0 15 data ASCII hex as returned b...

Страница 89: ...the instrument may be read by SETUP The same settings may then be stored by ending the same data back to the instrument As it sends all settings in a compressed format it is quicker than setting indi...

Страница 90: ...the resistance factor of an external current shunt to be divided into the measured voltage for a given input channel Format SHUNT channel resistance Arguments channel CH1 CH2 resistance shunt resista...

Страница 91: ...nt and dynamic response Format SMOOTH type dynamics Arguments type NONE NORMAL SLOW dynamics AUTO FIXED Reply none Example SMOOTH NORMAL FIXED SMOOTH NONE Notes It is not necessary to send both parame...

Страница 92: ...measurement speed Description Sets the minimum window size for the measurement Format SPEED value window Arguments value FAST MEDIUM SLOW VSLOW WINDOW Reply none Example SPEED SLOW SPEED WINDOW 0 1 N...

Страница 93: ...PPA25 26xx communications manual 2 81 START START Function Start datalog Description Initiate datalog data acquisition Format START Arguments none Reply none Example DATALOG RAM 0 02 START Notes...

Страница 94: ...ion of selected channel range number 1 16 range text overflow underflow status Format or STATUS STATUS channel STATUS channel Arguments channel CH1 CH6 Reply range number range text over under ok 1 16...

Страница 95: ...unications manual 2 83 STOP STOP Function Stop datalog Description Stop datalog data acquisition Format STOP Arguments none Reply none Example DATALOG RAM 0 02 START wait for data values STOP read dat...

Страница 96: ...USPEN SUSPEN Function Suspend data acquisition Description Disable the data acquisition to maximise the communication speed Format SUSPEN value Arguments value ON OFF Reply none Example FAST ON SUSPEN...

Страница 97: ...d offset for a temperature sensor connected to the torque input power transformer application mode Format TEMPER type scalefactor offset Arguments type DISABLED CENTIG FARHEN scale multiplying factor...

Страница 98: ...ual 2 86 TEMPER TEMPER Function Read the temperature Description Returns the measured temperature from a sensor connected to the torque input Format TEMPER Arguments none Reply single data value Examp...

Страница 99: ...t TORQSP type torquescaling speedscaling TORQSP OFFSET torqueoffset speedoffset Arguments type DISABLED ANALOG PULSED OFFSET Reply none Example TORQSP PULSED 10 50 speed measured by pulse torque scali...

Страница 100: ...l 2 88 TORQSP TORQSP Function Read the mechanical power Description Returns measured mechanical data values Format TORQSP Arguments none Reply 3 data values separated by commas power torque speed Exam...

Страница 101: ...s manual 2 89 USER USER Function Read the user data Description Returns up to 3 lines of user data Format USER Arguments none Reply 3 lines of ASCII terminated by CR Example USER Newtons4th Ltd R D de...

Страница 102: ...sign depending convention either negative if lagging current negative if leading current Format VARCON type Arguments type NEGLAG NEGLEA Reply none Example VARCON NEGLAG Notes An inductive load would...

Страница 103: ...ng with the details of the various parts of the instrument firmware Format VERSIO Arguments none Reply date code type cpu dsp fpga boot type 0 normal 30A 2 low current version 10A 4 high current versi...

Страница 104: ...26xx communications manual 2 92 VRMS VRMS Function Set up rms voltmeter Description Set mode to rms voltmeter Format VRMS Arguments none Reply none Examples VRMS Notes This has the same effect as MODE...

Страница 105: ...rmat VRMS phase results Arguments results RMS MEAN SURGE phase PHASE1 PHASE2 PHASE3 PHASES Reply RMS 6 data values separated by commas Vrms Arms Vdc Adc Vac Aac MEAN 6 data values separated by commas...

Страница 106: ...ion of SUM and neutral data Format WIRING type Arguments type SINGLE single ph 1 2PHASE 2 ph 2 wattmeter 3PH2WA 3 ph 2 wattmeter 3PH3WA 3 ph 3 wattmeter INDPH3 3 ph 2 wattmeter ph3 PHASE1 single ph 1...

Страница 107: ...or remove the zero Description Applies or removes a zero function depending on the measurement mode same as pressing ZERO key Resets the integration data and timer if in power integration mode Format...

Страница 108: ...display zoom level 1 2 single line display zoom level 2 data1 first data zoom level 1 or data for single line zoom level 2 data2 4 other data zoom level 1 data consists of line number for channel 1 o...

Страница 109: ...zoom level and data Format ZOOM Arguments Reply 5 integers separated by commas level 0 normal 1 2 4 value display zoom level 1 2 single line display zoom level 2 data1 4 zoom data data consists of li...

Страница 110: ...nd one query and return up to 64 different parameters from different phases in one response Step 1 Reset MULTILOG using the MULTIL 0 command This will clear any previously entered Multilog parameters...

Страница 111: ...f the power analyzers at N4L at present the PPA2500 PPA2600 have 78 possible functions Example extract from the Multilog function list Function Measurement Notes 1 frequency 2 watts 3 VA 4 VAr 5 power...

Страница 112: ...parameter 4 MULTIL 5 1 50 set Phase 1 RMS Voltage as parameter 5 MULTIL 6 2 50 set Phase 2 RMS Voltage as parameter 6 MULTIL 7 3 50 set Phase 3 RMS Voltage as parameter 7 Order parameter to be return...

Страница 113: ...end Multil query and read return string MULTIL returns a comma separated string as Example return string 5 0000E1 2 4500E2 2 4320E2 2 5421E2 1 0232E3 1 0152E3 1 0546E3 Frequency PH1 Watts PH2 Watts PH...

Страница 114: ...PPA25 26xx communications manual A 1 Appendix command summary COMMAND SUMMARY...

Страница 115: ...value ALARM1 type data high low ALARM2 type data high low APPLIC type setting BANDWI phase type BEEP BLANKI on off threshold CALVER string CALVER string CONFIG parameter data CONFIG parameter single...

Страница 116: ...hase R L C series R L C parallel tan Q LOWFRE on off MODE type MSLAVE type MULTILOG index phase func MULTILOG 1 30 floats as selected PFCONV convention PHASEM ratio PHASEM phase freq mag1 mag2 dB phas...

Страница 117: ...dow START STATUS channel range number range text over low ok STOP STREAM enable window STREAM disable STREAM data data data data data SUSPEN on off TEMPER type scale offset TEMPER single real data val...

Страница 118: ...eq CALFRQ Seven real data values CALIBR index value inputs CALIBR Single integer data value CALIDS string CALIDS String CALJIG value CALMOD value CALPHA index inputs CALRES CALSAV password CALSYS inde...

Страница 119: ...Power Analyser 3 Impedance Analyser 4 Power Integrator 5 Harmonic Analyser 7 Oscilloscope 3 Master slave Aux control 0 Disabled 1 Master 2 Slave 3 Sslave 4 Autozero manual or auto System options 0 Au...

Страница 120: ...w 2 Medium 3 Fast 4 Window 14 Smoothing Acquisition Control 0 Normal 1 Slow 2 None 15 Smoothing Response Acquisition Control 0 Auto reset 1 Fixed time 16 Baud rate Remote options RS232 0 19200 1 9600...

Страница 121: ...nternal 2 External Shunt 26 Input range channel 1 Range minimum range voltage 0 1v 1 3v 2 10v 3 30v 4 100v 5 300v 6 1Kv 7 3kV 27 Input range channel 2 Range minimum range current 0 100ma 1 300ma 2 1A...

Страница 122: ...1 Ranging voltage input attenuator ratio as required 35 External shunt channel 1 Ranging current input resistance value as required 38 Resolution remote options digit resolution 0 Normal 1 High 2 Bin...

Страница 123: ...splay 0 Current Frequency 1 Watts Current Frequency 2 VA Current Frequency 3 VAr Current Frequency 4 pf Current Frequency 5 Current Voltage Frequency 45 Function zoomed on 3 Main display 0 Watts Frequ...

Страница 124: ...ity Acquisition advance options 0 Normal 1 Narrow 50 Program 1 6 direct load System options 0 Disabled 1 Enabled 52 Frequency filter Acquisition control 0 Disabled 1 Enabled 53 Phase angle reference A...

Страница 125: ...rm3 term4 2 term1 term2 x term3 term4 3 term1 x term2 term3 term4 62 Argument term 1 Maths 0 Disabled 1 Constant 2 Voltage 3 Current 4 Torque 5 Speed 63 Sub argument term 1 For voltage and current arg...

Страница 126: ...PWM motor Drive 2 Lighting ballast 3 Inrush current 4 Power transformer 5 Standby power 71 Frequency filter Application options mode PWM Motor Drive 0 4KHz 1 1KHz 2 250Hz 72 Frequency tracking speed...

Страница 127: ...ue 77 Argument term 4 Maths 0 Disabled 1 Constant 2 Voltage 3 Current 4 Torque 5 Speed 78 Sub argument term 4 For voltage and current arguments only 0 rms 1 dc 2 ac 3 Fundamental 4 Peak 5 Mean 6 Ph Ph...

Страница 128: ...n options PWM motor drive 0 Disabled 1 Analogue speed 2 Pulsed speed 85 Torque scaling Nm V Enter Nm v value or Temperature v value 86 Speed scaling Hz V Enter rpm v value 87 Pulses per revolution Ent...

Страница 129: ...tion options mode PWM motor drive Nm offset value Application options mode Power transformer Temperature offset value 96 Speed offset Application options mode PWM motor drive rpm offset value 99 Compu...

Страница 130: ...ng edge 1 Falling edge 110 trigger type Scope 0 Auto 1 Normal 2 Single shot 111 trigger reference Scope 0 Voltage 1 Current 112 trigger phase Scope 0 Phase 1 1 Phase 2 2 Phase 3 113 cursors enable Sco...

Страница 131: ...ncy reference speed Enter ratio value 131 3 Phase 2 Wattmeter VA VAr Option Power Analyser 0 Low distortion 1 High Distortion 132 Integrator run time Hours Mode Power integrator enter figure 133 Integ...

Страница 132: ...1 1 ch1 ch2 152 RS232 printer enable Remote options 0 Disabled 1 Enabled 153 IEEE address Remote options GPIB mode enter address figures 154 Interface Remote options 0 RS232 1 LAN 2 GPIB 3 USB 155 Rec...

Страница 133: ...d 167 Alarm 2 data Alarm options 0 Zoom1 1 Zoom 2 2 Zoom 3 3 Zoom 4 168 Alarm 2 type Alarm options 0 Disabled 1 Linear 2 Alarm if high 3 Alarm if low 4 Outside window 5 Inside window 169 Alarm 2 high...

Страница 134: ...put ranging channel 3 Range autoranging voltage 0 Full Autorange 1 Autorange up 2 Manual 181 Input ranging channel 4 Range autoranging current 0 Full Autorange 1 Autorange up 2 Manual 182 Coupling pha...

Страница 135: ...ator 201 Enable channel 6 Range current input 0 Internal 2 External shunt 202 Input range channel 5 Range minimum range voltage 0 1v 1 3v 2 10v 3 30v 4 100v 5 300v 6 1Kv 7 3kV 203 Input range channel...

Страница 136: ...2MHz 1 Low dc 30KHz 2 dc dc 5Hz 208 Scale factor channel 5 voltage Ranging Enter figures as required 209 Scale factor channel 6 current Ranging Enter figures as required 210 External attenuator chann...

Страница 137: ...rated rms current integrator mode 23 average power factor integrator mode 24 integrated fundamental watts integrator mode 25 integrated fundamental VA integrator mode 26 integrated fundamental VAr int...

Страница 138: ...phase 55 current phase 56 harmonic voltage 57 harmonic current 58 dc voltage 59 dc current 60 ac voltage 61 ac current 62 peak voltage 63 peak current 64 voltage crest factor 65 current crest factor 6...

Страница 139: ...84 voltage rms deviation transformer mode 85 voltage fundamental deviation transformer mode 86 voltage phase deviation transformer mode 87 99 reserved for future expansion Note Functions 78 and 81 are...

Страница 140: ...mbers available these Special functions re use function numbers that apply to other measurements for Phases 1 to 3 Examples for setting up individual special function measurements multil 0 Setting to...

Страница 141: ...230 1066 international 44 116 230 1066 Fax 0116 230 1061 international 44 116 230 1061 E mail address sales newtons4th com office newtons4th com web site www newtons4th com At Newtons4th Ltd we have a...