Description and Speci cations
2.2.
Speci cations
2.2.1.
Optical Parameters
Base unit
PSCOUT2-BASE
Wavelength range
420 - 1600 nm
PSCOUT2-BLUE-PD
PSCOUT2-RED-PD
PSCOUT2-NIR-PD
PSCOUT2-IR-PD
Wavelength range
420 - 550 nm
520 - 750 nm
700 - 1100 nm
1100 - 1600 nm
PSCOUT2-BLUE-PMT
PSCOUT2-RED-PMT
PSCOUT2-NIR-PMT
PSCOUT2-IR-PMT
Wavelength range
420 - 550 nm
520 - 750 nm
700 - 1100 nm
1100 - 1600 nm
Suggested detector type
Depending on laser repetition rate:
PD < 1 MHz and PMT > 1 MHz repetition rate
Scan range
150 fs - 15 ps
(150 fs, 500 fs, 1.5 ps, 5 ps, 15 ps)
Measureable pulse width
20 - 3500 fs (selectable ranges)
For all detector types
Min. input power
> 2 mW
Max. input power
< 250 mW or < 1
µ
J
(which ever results in the lower average power)
Repetition rate
300 Hz to 10 GHz,
of input laser
PMT detector suggested above 1 MHz
Linearity
1 % of actual scan range
Polarization
Linear / horizontal
Beam characteristics
diameter > 2 mm
1
/e
2
PC interface
USB 2.0
Remote control
TCP/IP (SCPI Standard) via Control Software
Trigger Input
level
0.1 ... 5 Vrms 50
Ω
0.1 ... 8 Vpp 1 k
Ω
impedance 50
Ω
/ 1 k
Ω
repetition rate 300 Hz ... 50 kHz
width > 50 ns
Dimensions
Table 2.1.: Speci cations
Pulse
Scout2 Version 1.1
12