SN
set axis displacement units
IMM
PGM
MIP
USAGE
♦
♦
SYNTAX
xx
SN
nn or xx
SN
?
PARAMETERS
Description
xx
[ int ] -
axis number
nn
[ int ] -
displacement units
Range
xx
-
1
to
MAX AXES
nn
-
0
to
10
where
0 = encoder count
6 = micro-inches
1
=
motor
step
7
=
degree
2
=
millimeter
8
=
gradient
3
=
micrometer
9
=
radian
4
=
inches 10
=
milliradian
5
=
milli-inches
11
=
microradian
or
?
to read present setting
Units
xx
-
none
nn
-
none
Defaults
xx
missing:
error 37, AXIS NUMBER MISSING
out of range:
error 9, AXIS NUMBER OUT OF RANGE
nn
missing:
error 38, COMMAND PARAMETER MISSING
out of range:
error xx2, PARAMETER OUT OF RANGE
DESCRIPTION
This command is used to set the displacement units for the for axis
xx
.
Note:
The unit of measure as used with this controller is intended as a label only. It is
the user’s responsibility to convert and resend all affected parameters (e.g.,
velocity, acceleration, etc…) when switching from one unit of measure to
another.
RETURNS
If the “?” sign takes the place of
nn
value, this command reports the current
setting
REL. COMMANDS
FR
-
set full-step resolution
SU
-
set encoder resolution
EXAMPLE
2SN
|
read displacement unit setting of axis # 2
2
|
controller returns a
value 2 (millimeter) for axis #2
2SN 0
|
set displacement unit to 0 (encoder count) for axis #2
Section 3 – Remote Mode
3-133
Содержание ESP300 Series
Страница 1: ...ESP300 Motion Controller Driver User s Manual...
Страница 4: ...ESP300 1999 EU Declaration of Conformity iv Preface...
Страница 29: ...This page is intentionally left blank Section 2 Modes of Operation 1 15...
Страница 30: ......
Страница 230: ...4 16 Section 4 Advanced Capabilities...
Страница 274: ...This page is intentionally left blank 6 6 Section 6 Servo Tuning...
Страница 290: ...This page is intentionally left blank B 6 Appendix B Trouble Shooting and Maintenance...