QR
reduce motor torque
IMM PGM MIP
USAGE
♦
♦
♦
SYNTAX
xx
QR
nn1,nn2 or xx
QR
?
PARAMETERS
Description
xx
[ int ]
-
axis number
nn1
[ int ]
-
delay period
nn2
[float ]
-
motor current reduction percentage
Range
xx
-
1
to
MAX AXES
nn1
-
0
to
60000
nn2
-
0
to
100
Units
xx
-
none
nn1
-
milliseconds
nn2
-
percent of max. motor current
Defaults xx
missing:
error 37, AXIS NUMBER MISSING
missing parameter:
error 38, COMMAND PARAMETER MISSING
out of range:
error 9, AXIS NUMBER OUT OF RANGE
DESCRIPTION
This command automatically reduces the specified step motor’s current (i.e.,
torque) output to the requested percentage
nn2
after motion has stopped and the
specified time
nn1
has expired. The purpose of this command is to help reduce
the motor heating typically generated by stepper motors. If
xx
is equal to 0, the
torque reduction parameters get applied to all axes.
Note:
This command does not affect DC servo motors and pulse stepper motors.
Note:
Time parameter
nn1
is only effective for ESP300 motion controller
RETURNS
If “?” sign is issued along with command, the controller returns the torque
reduction settings for the specified axis.
REL. COMMANDS
QM
-
set motor type
QI -
et motor current
EXAMPLE
2QR1000,50
|
reduce motor #2 torque to 50%, 1000 msec after a move
done
2QR?
|
query motor #2 torque reduction settings
1000,50
|
controller returns 1000 msec and 50%
3-118
Section
3
–
Remote
Mode
Содержание ESP300 Series
Страница 1: ...ESP300 Motion Controller Driver User s Manual...
Страница 4: ...ESP300 1999 EU Declaration of Conformity iv Preface...
Страница 29: ...This page is intentionally left blank Section 2 Modes of Operation 1 15...
Страница 30: ......
Страница 230: ...4 16 Section 4 Advanced Capabilities...
Страница 274: ...This page is intentionally left blank 6 6 Section 6 Servo Tuning...
Страница 290: ...This page is intentionally left blank B 6 Appendix B Trouble Shooting and Maintenance...