MD
read motion done status
IMM
PGM
MIP
USAGE
♦
♦
SYNTAX
xx
MD?
PARAMETERS
Description
xx
[ int ] -
axis number
Range
xx
-
1
to
MAX AXES
Units
xx
-
none
Defaults
xx
missing:
error 37, AXIS NUMBER MISSING
out of range:
error 9, AXIS NUMBER OUT OF RANGE
DESCRIPTION
This command is used to read the motion status for the specified axis
n
. The MD
command can be used to monitor Homing, absolute, and relative displacement
move completion status.
RETURNS
nn
-
0
or
1
where 0 = motion not done (FALSE)
1
=
motion
done
(TRUE)
REL. COMMANDS
PA
-
move to an absolute position
PR
-
move to a relative position
3PR2.2
|
start a relative motion of 2.2 on axis # 3
OR
-
move to home position
EXAMPLE
3MD?
|
read axis #3 move done status
1
|
controller returns status 1 (motion done) for axis # 3
3MD?
|
read axis #3 move done status
0
|
controller returns status 0 (motion not done) for axis # 3