45
Maintenance
Annual Inspection
Contact New Cosmos or your New Cosmos representative to perform a periodic inspection
including sensor calibration and filter replacement at least once a year to maintain the
product accuracy.
Cleaning
If the detector is dirty, wipe it off with a soft dry cloth, or a cloth fully wrung out of water (dry
enough not to leave the product surface wet). Do not use any alcohol or detergent.
Consumable Parts
Part Name
Model
Remarks
Filter element for drain filter
FE-2
Replace at least once a year or when
dirty or wet.
Filter element for cooling
drain filter
FE-10
Replace at least once a year or when
dirty or wet.
Sensor spacer
with two gaskets
Replace when the sensor spacer or its
gasket is damaged or deformed.
Elastomer case
Replace when worn or damaged.
1m gas sampling tube
SH-301K-1A
With drain filter and probe nozzle
included
Drain filter
DF-112
Attachment
AT-2B
Short probe
Oxygen sensor
OS-5
For replacement
Nickel metal hydride AA
battery (rechargeable)
HR-3UTG
Replace when the continuous
operation time becomes extremely
short or every two years at least.
*The warranty period for consumable parts is one year from the date of purchase. The
warranty terms and conditions are the same as the one for the product (gas detector).
The recommended replacement cycle for combustible sensors is
three years. Replace the sensor with a new one every three years
to ensure correct detection.
The recommended replacement cycle for oxygen sensors is two
years. Replace the sensor with a new one every two years to
ensure correct detection.
The above-recommended cycles are only an estimate based on
normal use and proper maintenance without exposure to high
concentration gas or gas poisoning; therefore, no guarantee is
provided.
The replacement cycle changes depending on the use and
environment conditions. Check your detector before use for
normal operation, when it was used outside the specified
conditions, such as high temperature/humidity, impact by falling
from a high place, exposure to water splash, high concentration
gas or gas poisoning.