24
The Ready, Stimulation, and Rest phases are the same as Stroke Rehab mode.
The only difference is that for a given signal level detected in the Ready or
Rest modes, the vertical column will be higher in SCI mode as compared to
Stroke Rehab mode.
Manual Mode
This mode is similar to Stroke Rehabilitation mode, but requires manual
setting of the threshold. Set the threshold in the programming mode as
described in the section “Programming the NeuroMove®” above. In Manual
mode, the threshold does not automatically adjust up or down as in Stroke
Rehab mode. The Ready, Stimulation, and Rest phases are the same as Stroke
Rehab mode.
Stimulation-Only mode
In this mode, the unit acts as a neuro-muscular stimulator only. I.e., it does
not need to be triggered to produce stimulation. In effect it continuously
repeats the Stimulation and Rest phases of the normal therapy cycle. There
is not a Ready phase since signal levels from the target muscle group are not
monitored.
Stim Intensity:
15 mA
To increase the stimulation intensity, press the Up button. To decrease, press
the Down button.
Audio Prompts
The NeuroMove® provides an array of audio prompts to assist the user in the
operation of the machine. Audio prompts can be enabled or disabled in the
programming mode. See Programming Section above.