17
Ramp Off Time:
1 sec.
To increase the ramp off time, press the Up button. To decrease, press the
Down button.
Sensitivity Range
To change the sensitivity range, press the program button once again.
The sensitivity range describes the range of signal levels from the patient that
the NeuroMove® can detect.
When in Stroke Rehab operational mode there is only one range,
1.25 – 1000
µV (microvolts). Switch to SCI operational mode if Stroke Rehab is not
sensitive enough (displayed signal levels are at a very low level or bottomed out).
In SCI mode, two ranges are selectable: 0.25 - 25 µV and 0.50 - 50 µV. Use the
latter if the former is too sensitive (levels are reaching maximum).
Sensitivity:
1.25-1000 uV
Press the Up or Down key to select the desired range.