15
Pulse Width
To change the stimulation pulse width, press the program button once again.
The pulse width can be selected in the range from 50 - 400 µs (microseconds)
in 5 microsecond increments. The default is 200 microseconds. This
parameter normally does not need to be changed by the user.
Pulse Width:
200 uS
To increase the pulse width, press the Up button. To decrease, press the Down
button.
Stimulation Period
To change the stimulation period, press the program button once again.
The stimulation period is the time during which unit delivers electrical energy
to the patient to create a muscle contraction. The stimulation time can be
selected in the range from 0.5 - 30 seconds. The default is 5 seconds.
Note: the word “Stim” along with the intensity appears in the upper left
corner of the display during the stimulation period.
Stim. Period:
5 sec.
To increase the stimulation period, press the Up button. To decrease, press
the Down button.