Neomeris 850909 Скачать руководство пользователя страница 1

User Manual 

 

Neomeris Basic 

pH/Temperature  

Pocket-Tester 

 

pH-/°C-/°F-Meter in  

pocket size 

 
 

 

 

 

 

 

www.neomeris.de 

 

 
Automatic Temperature 

      Measuring  unit 

Compansation Indicator 

     on main LCD 

Instability Indicator 

     Main LCD 

Calibration Indicator 

     Temperature 

Low Battery Indicator 

     reading  

 

     Secondary LCD 

 

Operational Guide 

 

• Remove the electrode cap and immerse the 
meter in 

Neomeris KCL electrolyte solution

 

(890707) for 2 hours to activate the electrode. 
 
• Turn the meter on by pressing the ON/OFF 
button. All the used segments on the LCD will be 
visible for 1 second or as long as the button is 
pressed. 
 
• Immerse the electrode in the solution to be 
tested. Stir gently and wait for the reading to 
stabilize, i.e. the instability indicator (hourglass) 
on the LCD turns off. 
 
• The pH value is automatically compensated for 
temperature and will be displayed on the main 
LCD, while the temperature is shown on the 
lower part of the LCD. 
 
• 

To freeze the display

, while in measurement 

mode, press and hold the SET/HOLD button. 
The “HOLD” message appears on the secondary 
display and the reading will be frozen on the 
LCD. Press any button to return to normal mode. 
 
• 

To turn the meter off

, press the ON/OFF 

button. The “OFF” message will appear on the 
secondary display. Release the button.

 

 

Note: 
 

• Before taking any measurement, make sure the 
meter is calibrated (the CAL tag is on). 
• Always switch off the meter after use. Rinse the 
electrode with water to minimise contamination 
and store it in the protective cap with a few drops 
of KCL (

890707

) or pH7 (

890688

) solution.  

 

NEVER USE DISTILLED OR DEIONIZED 
WATER FOR STORAGE PURPOSE 

Calibration Procedure 

 

For better accuracy, frequent calibration of the 
tester is recommended. Calibration is also 
necessary after electrode replacement, after 
testing aggressive chemicals and where extreme 
accuracy is required. 
 
• From normal operation mode, press and hold 
the ON/OFF/CAL button until the “OFF” message 
on the secondary LCD is substituted by “CAL”. 
Release the button. 
 
• The instrument enters the calibration mode by 
displaying “pH 7.01 USE” (or “pH 6.86 USE” if 
the NIST buffer set was selected). 
 
• 

For a

 

single-point calibration

, immerse the 

electrode in any buffer, i.e. pH 4.01, 7.01 (or 
6.86), 10.01 (or 9.18). 
 
• The meter activates the automatic buffer 
recognition. If no valid buffer is detected, the 
meter keeps the USE indication active for 12 
seconds, and then replaces it with WRNG, 
indicating that the sample being measured is not 
a valid buffer. Otherwise, if a valid buffer is 
detected, then its value is shown on the primary 
display, and REC appears on the secondary 
LCD. 
 
• If the pH 7.01 (or pH 6.86) was used, press the 
SET button to exit the Calibration mode and the 
“OK 1” message will appear on the display. The 
first calibration point is stored and the meter 
returns to normal measurement mode. 
 

It is always recommended to perform a  
2-point calibration to obtain more accurate 
measurement results. 

 
• 

For a two-point calibration

, immerse the 

electrode in pH 7.01 (or pH 6.86) buffer solution. 
 

• After the first point has been accepted, the 
meter will then ask for the second buffer and the 
message „pH 4.01 USE” will appear. 
 
• Rinse the electrode and immerse it in the 
second solution (pH 4.01, 10.01 or 9.18). 
 
• If a valid buffer value is detected, the REC 
message is displayed and the meter completes 
the calibration procedure. The LCD shows the 
accepted value with the „OK 2” message and the 
instrument returns to the normal measurement 
mode. Otherwise, if no valid buffer is detected, 
the meter displays the WRNG message 
 

Note:

 When the calibration procedure is 

completed, the CAL tag is turned on. 
 
• 

To quit the procedure

 

and return to the last 

calibration data, after entering the calibration 
mode press the ON/OFF button. The secondary 
LCD displays “ESC” for 1 second and then the 
meter returns to the normal measurement mode. 
 
• 

To reset to the default values

 

and clear a 

previous calibration, press the SET/HOLD button 
after entering the calibration mode and before 
the first point is accepted. The secondary LCD 
displays “CLR” for 1 second, the meter resets to 
the default calibration and the CAL tag on the 
LCD turns off. 

 

Setup

 

 

Setup mode allows the selection of temperature 
(°C or °F) and pH buffer set for calibration. 
To enter the Setup mode, press the ON/OFF 
button until “CAL” on the secondary LCD is 
replaced by “TEMP” and the current temperature 
unit (e.g. TEMP °C). Then: 
 

 

for °C/°F selection:

 use the SET/HOLD button; 

then press the ON/OFF button once to enter the 
buffer set selection or twice to return to the 
normal measurement mode. 
 