GravityPro
Oral Syringe with Vent Zone and Integrated Hanger
For Oral / Enteral Use
Instructions for Use
Gravity Feed Features
1. Designed with an integrated hanger for user convenience.
2. Incorporates a vent zone that allows the plunger to stay in place.
3. Minimizes air contact of nutritional fluids
4. Promotes Aseptic Technique
5. Allows air flushing with the same syringe
6. May reduce spillage of nutritional fluids
Other Features
1. Pump compatible
2. This product is not made with natural rubber latex, DEHP, or BPA.
Indications for Use
The device is indicated for use as a dispenser, a measuring device
and an oral fluid transfer device. It is used to inject fluids into
the body
via extension sets and feeding tubes in neonatal and pediatric patients.
Contraindications
The GravityPro Oral Syringe is intended for one-time use.
Each syringe should be used for one dispensing and one patient.
Warnings
•
Do not autoclave or re-use syringe as this may increase
risk of leakage or biological infection.
•
Tip caps are a choking hazard. Keep out of reach of children.
Precautions
•
Always attach the syringe to a NeoMed feeding tube or extension
set prior to moving the plunger gasket into the vent zone, as fluid will
flow immediately when the plunger gasket is positioned in the vent zone.
•
Do not invert the syringe when the plunger gasket is within the vent zone.
Feeding Preparation
1. Follow physician’s instructions or hospital protocol in opening package and use of contents. Use aseptic technique.
2. Identify the vent zone on the syringe (See Figure 1)
3. Fully depress the plunger before filling the syringe. Draw liquid (e.g. formula or mother’s milk) into syringe. For
accurate meas-urement, use the black leading edge of the gasket toward the plunger tip. The accuracy of
measurements is ±4% of expelled volume. Do not fill or attempt delivery past the maximum graduation. Do not
position plunger gasket in the vent zone during preparation.
4. To expel excess air bubbles, ensure the plunger is not positioned in the vent zone. Hold syringe with tip pointing
up and tap side of syringe to cause bubbles to rise toward tip. Draw back slightly on plunger taking care to not
draw back into the vent zone. Push plunger upward to eject air in barrel and tip. Repeat until bubbles are eliminated.
5. Place the orange tip cap on the syringe and label as per hospital protocol (label not included).
NeoMed and the NeoMed logo are registered trademarks of NeoMed, Inc.
© 2017 NeoMed, Inc.
LBL-OV-003 Rev 2
Effective 04-2017
Figure 1: Vent Zone
How Supplied:
20 mL
REF
Size
Pouch
Qty
Case
Qty
100
40
100 mL
1
1
NM-S20EOV
NM-S100EOV