GravityPro

Oral Syringe with Vent Zone and Integrated Hanger

For Oral / Enteral Use

Instructions for Use

Gravity Feed Features

1. Designed with an integrated hanger for user convenience.
2. Incorporates a vent zone that allows the plunger to stay in place.
3. Minimizes air contact of nutritional fluids
4. Promotes Aseptic Technique
5. Allows air flushing with the same syringe
6. May reduce spillage of nutritional fluids

Other Features

1. Pump compatible
2. This product is not made with natural rubber latex, DEHP, or BPA.

Indications for Use

The device is indicated for use as a dispenser, a measuring device 
and an oral fluid transfer device. It is used to inject fluids into

 

the body

via extension sets and feeding tubes in neonatal and pediatric patients.

Contraindications

The GravityPro  Oral Syringe is intended for one-time use. 
Each syringe should be used for one dispensing and one patient.

Warnings

• 

Do not autoclave or re-use syringe as this may increase 

   risk of leakage or biological infection.

• 

Tip caps are a choking hazard. Keep out of reach of children.

Precautions

• 

Always attach the syringe to a NeoMed feeding tube or extension 

   set prior to moving the plunger gasket into the vent zone, as fluid will
   flow immediately when the plunger gasket is positioned in the vent zone.

• 

Do not invert the syringe when the plunger gasket is within the vent zone.

Feeding Preparation

1. Follow physician’s instructions or hospital protocol in opening package and use of contents. Use aseptic technique.
2. Identify the vent zone on the syringe (See Figure 1)
3. Fully depress the plunger before filling the syringe. Draw liquid (e.g. formula or mother’s milk) into syringe. For 
    accurate meas-urement, use the black leading edge of the gasket toward the plunger tip. The accuracy of 
    measurements is ±4% of expelled volume. Do not fill or attempt delivery past the maximum graduation. Do not 
    position plunger gasket in the vent zone during preparation.
4. To expel excess air bubbles, ensure the plunger is not positioned in the vent zone. Hold syringe with tip pointing 
    up and tap side of syringe to cause bubbles to rise toward tip. Draw back slightly on plunger taking care to not 
    draw back into the vent zone. Push plunger upward to eject air in barrel and tip. Repeat until bubbles are eliminated.
5. Place the orange tip cap on the syringe and label as per hospital protocol (label not included).

NeoMed and the NeoMed logo are registered trademarks of NeoMed, Inc.

© 2017 NeoMed, Inc.

LBL-OV-003 Rev 2

Effective 04-2017

Figure 1: Vent Zone

How Supplied:

20 mL

REF

Size

Pouch

Qty

Case

Qty

100

40

100 mL

1

1

NM-S20EOV

NM-S100EOV