Using the N-560
N-560 Operator’s Manual
39
Press the
Adjust Up
button or
Adjust Down
button to select the desired
baud rate. Possible settings are:
•
24 (2,400 bps)
•
96 (9,600 bps)
•
192 default (19,200 bps).
Setting the Data Port Protocol
With the N-560 in the normal monitoring mode:
1. Simultaneously, press the
SpO2 Alarm Limit
button and
Pulse Rate
Alarm Limit
button for three seconds. Menu option 1 will be
displayed.
2. Press the
Adjust Up
button until menu item option 5 is displayed.