Performance Verification
10 N5500 Service Manual
Performance
Verifi
cation, N5500
3.3 Performance Tests
The battery charge and battery performance test should be performed before monitor
repairs whenever the battery is suspected as being a source of the problems. All other
tests may be used following repairs or during routine maintenance (if required by your
local institution). Before performing the battery performance test, ensure that the battery
is fully charged. This section is written using Nellcor factory-set power-up defaults. If
your institution has preconfigured custom defaults, those values will be displayed.
3.3.1 Battery Charge
1. Connect monitor to AC power source using proper power cord.
2. Verify AC Power indicator is lit.
3. Charge battery fully for at least 18 hours.
4. The only way to check for a full charge is to perform the procedure in paragraph
3.3.2 “Battery Performance Test.”
Note: The battery may require a complete charge/discharge cycle to restore its normal
capacity, depending on its previous usage.
3.3.2 Battery Performance Test
1. The N5500 monitor is specified to typically operate on battery power a minimum of
1 hour, at 25°C, with no printing, and one NIBP measurement every 5 minutes.
Before performing this test, ensure that the battery is fully charged (paragraph 3.3.1).
2. Ensure monitor is not connected to AC power.
3. With N5500 turned off, press Power On/Off switch and verify battery icon appears at
bottom of display after power-on self-test is completed. Boxes in battery icon should
all be filled, indicating battery is charged.
4. Connect
Nellcor
SRC-MAX SpO
2
simulator to monitor via DOC-10 sensor cable.
5. Connect NIBP simulator to monitor via Colin-NO.EM hose.
6. Set SRC-MAX as follows: SpO
2
of 75% and pulse rate of 60 bpm.
7. Set NIBP simulator to simulate pressure setting of 120/80 mmHg and heart rate of 80
bpm.
8. Verify monitor is responding to SpO
2
simulator signal and audible alarm is sounding.
Use knob to select SpO
2
Menu and silence SpO
2
audible alarm (SpO
2
alarm suspend
condition).