Specifications
N-550
87
Safety Tests
Ground Integrity
100 milliohms or less
Earth Leakage Current
Enclosure Leakage Current
AC Line
Polarity
AC Line
Cord
Neutral
Line Cord
IEC 60601
-
1,
AAMI/
ANSI
-
ES1
UL 2601
-
1
Normal
Closed
Closed
500
µ
A
300 µA
Reversed
Closed
Closed
500
µ
A
300 µA
Normal
Open
Closed
1000
µ
A
500 µA
Normal
Closed
Open
1000
µ
A
500 µA
AC Line
Polarity
Neutral Line
Cord
Power Line
Ground
Cord
IEC 60601
-
1
UL 2601
-
1,
AAMI/
ANSI
-
ES1
Normal
Closed
Closed
100
µ
A
100
µ
A
Normal
Closed
Open
500
µ
A
300
µ
A*
Normal
Open
Closed
500
µ
A
300
µ
A
Reversed
Closed
Closed
100
µ
A
100
µ
A
Reversed
Open
Closed
500
µ
A
300
µ
A*
Reversed
Closed
Open
500
µ
A
300
µ
A
* = AAMI/ANSI
-
ES1 does not include opening line conductor.
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...