(3)  Rotate the Filter Selection Lever to choose between the 3 available filters below: 

Marker 

Filter 

Use 

White 

UV 

Normal use. Filter will not alter the color tone 

Green 

Red-free 

Blood will appear dark, facilitating recognition of blood 
vessels or hemorrhages. 

Blue 

Cobalt Blue 

Used in conjunction with fluorescein 

The color of the Battery Lamp will change according to the amount of battery power 
remaining. The Battery Lamp will turn orange to indicate that there is less than 20 minutes 
remaining (at maximum illumination level). 

Color 

Condition of Battery Pack and Necessary Actions 

Green 

Battery operational. Continue use 

Orange 

Little battery power remaining. Charge immediately 

Unlit 

Battery fully discharged. Charge immediately 

3.2.5  Examination 

 

WARNING 

Do not illuminate the patient’s eye for more than 1 minute per eye and be sure 
to keep illumination at minimal intensity needed to view the fundus. 

(1)  Holding an aspherical lens in one hand, direct the illumination into the patient eye. It 

is recommended to place the unoccupied fingers of the hand holding the aspherical 
lens on the patient’s face for stability. 

(2)  Bring the aspherical lens close to the patient eye and gradually move away from the 

eye until a clear image of the fundus is obtained. 

*Note:  The distance between the patient eye and aspherical lens is dependent on 

the optical power (diopter, 1 D = 1 / focal distance in meters) of the aspherical 
lens. 

Optical Power 

Description of Image 

14 D to 16 D 

High magnification, bright image, narrow field of view 

20 D 

Typically used in standard examinations 

28 D to 30 D 

Low magnification, dim image, wide field of view 

The illumination system and observation system can be configurated as follows to 
optimize examination 

<Examining Large Pupils> 

When examining the fundus through large pupils, 
raise both the Illumination Angle Adjustment Lever 
and Observation Angle Adjustment Lever. The 
illumination axis and observation axes are 
distanced, allowing for better stereoscopic vision 
with reduced glare (see Fig. 3.9a). 

Illumination 

Light

 

Observation 

Light

 

Fig. 3.9a 

Содержание IO-A LED

Страница 1: ...is is a battery powered device to examine the inside and fundus of the eyes by illuminating with an LED bulb The device is powered by the Battery Pack IO BP which can be attached to the Headband or a...

Страница 2: ...ng the IO BP to the Scope Unit 11 3 2 Operation 12 3 2 1 Wearing the IO 12 3 2 2 Positioning the Scope Unit 12 3 2 3 PD Adjustment 12 3 2 4 Illumination 12 3 2 5 Examination 13 3 2 6 Using the Teachin...

Страница 3: ...and caution you of potential hazards to yourself and others Potentially hazardous situation which if not avoided could result in death or serious injury Potentially hazardous situation which if not av...

Страница 4: ...ses Catalog number product code Serial number 1 3 Safety Information For your safety please comply with the following precautions Do not use battery packs or AC adaptors other than the standard access...

Страница 5: ...ing Failure to do so may result in electric shock Take care to only touch insulated parts when inserting the power plug into the power outlet Failure to do so may cause electric shock or fire Do not s...

Страница 6: ...ned off during transport or storage and only use the specified casing Failure to do so may result in deformation device failure or fire When disconnecting cables be sure to hold the plug Never pull th...

Страница 7: ...do not use the IO and contact your local dealer immediately Set Type Item Set 4 Set 3 Set 2 IO Main Unit See 2 2 1 1 1 1 IO BP See 2 2 2 1 1 1 AC Adapter example 1 1 1 Extension Cord 1 m 1 1 1 Belt Ho...

Страница 8: ...nation field 7 Aperture Selection Marker 8 Eyepiece A 2 D diopters lens is included 9 PD Adjuster Adjusts the pupil distance PD The scale 54 mm to 74 mm is indicated above the Eyepiece 10 Illumination...

Страница 9: ...ds the Scope Unit Can be flipped up 17 Overband Fixation Knob Loosen to flip up the Overband Tighten to keep the Overband in place 18 Headband Cover s Covers the adjustment mechanics of the Headband F...

Страница 10: ...ff and allows for continuous intensity adjustment 4 Illumination Off Mark 5 Illumination Intensity Mark 6 Battery Lamp Indicates the remaining battery capacity 7 Indicates power output from the Lamp J...

Страница 11: ...ange 3 Charging time is approximately 2 5 hours from a fully discharged state Note The IO cannot be illuminated while the DC Plug is connected 4 Once fully charged the CHARGE Lamp will light green 5 U...

Страница 12: ...th a click see Fig 3 4 Note The plug and jack interlock at a single configuration Rotate the plug as needed to connect smoothly 2 When using the Belt Hook the Lamp Cord attached to the Scope Unit is n...

Страница 13: ...n the Scope Unit Fixation Knob to fix in the position with the best view when observed through the eyepieces see Fig 3 7 3 2 3 PD Adjustment 1 Turn the Observation Angle Adjustment Lever upward to ful...

Страница 14: ...lens in one hand direct the illumination into the patient eye It is recommended to place the unoccupied fingers of the hand holding the aspherical lens on the patient s face for stability 2 Bring the...

Страница 15: ...observation axes remain apart see Fig 3 9c Examining Vertically Elliptical Pupils When examining the fundus through vertically elliptical pupils raise the Illumination Angle Adjustment Lever and lowe...

Страница 16: ...detaching the Teaching Mirror from the Scope Unit depress the mark on both the upper and lower faces of the Teaching Mirror and pull gently Note Do not attempt to detach the Teaching Mirror forcefull...

Страница 17: ...pe the surface using a dry lint free cloth or cloth moistened with diluted neutral detergent and then wipe dry Use a cloth moistened with alcohol to disinfect 4 2 Replacing the LED Bulb As LED bulbs h...

Страница 18: ...operate If such events should occur contact your local dealer to purchase a new Battery Pack 4 4 Disposal Dispose of the IO and its accessories excluding the IO BP in accordance with your local regula...

Страница 19: ...oned Move the lever s until you feel a click Observation image blurry Foreign matter is adhered to the eyepiece or observation system Clean in accordance with 4 1 1 Foreign matter is adhered to the as...

Страница 20: ...tion Degree of protection against electric shock Internally powered ME equipment Applied parts No applied parts Degree of protection against harmful ingress of water or particulate matter IPX0 Method...

Страница 21: ...ort contact your local dealer or Neitz located at the following address LOCAL DEALER Neitz Instruments Co Ltd 4F Ichibancho Court 15 21 Ichibancho Chiyoda ku Tokyo 102 0082 Japan Phone 81 3 3237 0552...

Страница 22: ...Ophthalmic instruments Fundamental requirements and test methods Part 2 Light hazard protection indicates that the IO is classified to be Group 2 and ophthalmic instrument for which a potential light...

Страница 23: ...O is intended for use in the electromagnetic environment specified below The customer or the user of the IO should assure that it is used in such an environment Emissions Test Compliance Electromagnet...

Страница 24: ...000 4 5 1 kV line s to line s 1 kV line s to line s Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power...

Страница 25: ...d RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with t...

Страница 26: ...cy of transmitter m 150 kHz to 80 MHz 1 2 80 MHz to 800 MHz 1 2 800 MHz to 2 5 GHz 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maxi...

Страница 27: ...25 25 Notes...

Страница 28: ...Japan Phone 81 3 3237 0552 FAX 81 3 3237 0554 E mail int neitz co jp URL http www neitz ophthalmic com Distributed by As part of our policy of continued product improvement we reserve the right to alt...

Отзывы: