Performing the inhalation process |
3
Version: NTI_HM-200_IFU-EN_G
Page
18
3
Performing the inhalation process
Programs of the SaHoMa
®
-II control unit
The SaHoMa
®
-II control unit is equipped with two programs:
Program 1: Automatic switch-off
Program 1
is a program that automatically switches off the SaHoMa
®
-II control unit as
soon as there is no more inhalation solution in the medication cup of the SaHoMa
®
-II
mesh nebulization unit.
Program 2: Time program
Program 2
should be used for medications with increased viscosity or those that tend
to foam, as these medications cause a premature switch-off of the
SaHoMa
®
-II mesh nebulization unit in Program 1.
Program 2 is a time program in which you can set the desired nebulization time. The
SaHoMa
®
-II control unit switches off after the set time has elapsed. An automatic
switch-off, when the medication is used up, does not occur in program 2.
In
chapter
3.2 „Switching on the SaHoMa
®
-II control unit in program 1“
and
chapter 3.3 „
Switching on the SaHoMa
®
-II control unit in program 2 “
the use of
both programs is explained in detail.