
Quality Messages and Quality Grades
84
EasyOne Pro™ / EasyOne Pro™ LAB
5
Quality Messages and Quality Grades
5.1 FVC/FVL Quality Messages and Quality Grades
Some Basic Facts
End-of-Test criteria, quality criteria and quality grading are based upon the
published standards [1], [4], [11], [15] (
Quality grading is based on [4], [11], [15] (
Additional remarks
The main articles [2] and [3] do not numerically define the minimum
expiratory peak flow time (PEFT) that is required for an acceptable test. In
this case, 160 ms is used.
The end-of-test criteria for FVC tests are as follows: A test ends when the
volume change during the last 2 seconds is <45 ml, or an inspiratory
volume >150 ml is detected. This end-of-test criterion slightly differs from
the published criterion in [3] which is 25 ml in the last second. Reason:
When the 25 ml in 1 second criterion is applied to the 24 waveforms also
defined in [1] and [2], several curves will end too early, and these tests
might fail. We therefore decided to slightly change the criterion to 45 ml
in 2 seconds.
In the following tables the quality messages and the quality criteria are
detailed.
The end-of-test criterion for an FVC test is as follows: A test ends when the
volume change during the last 2 seconds is <45 ml, or an inspiratory volume
>150 ml is detected. When this end-of-test criterion is met, the following
quality messages are checked:
Содержание easyone pro
Страница 1: ...Version V04b EasyOne Pro EasyOne Pro LAB Operator s Manual ...
Страница 133: ...EasyOne Pro EasyOne Pro LAB 133 Specifications Power con sumption max 80 VA Life time 7 years ...
Страница 138: ...Appendix 138 EasyOne Pro EasyOne Pro LAB Fig 12 1 ...
Страница 184: ...Appendix 184 EasyOne Pro EasyOne Pro LAB ...