Danger
Explosion Hazard—
Ensure that persons working with oxygen have undergone
special training and are aware of the particular properties
of oxygen to guarantee safe handling of oxygen without
accidents. Furthermore, these persons must be familiar
with this operator manual and must have understood its
content.
Do not use hand cream or other skin care products.
Oxygen intensively promotes combustion; therefore,
flammable substances must be kept away from oxygen.
Unintended release of oxygen in confined spaces
increases risk of fire; smoking and open flames are
prohibited. Possible sources of ignition must be
eliminated. Rooms must be well ventilated.
The oxygen inlet must periodically be checked for leaks
(at least once a month).
Use only fittings approved for use with oxygen; these
fittings must be properly connected. All fittings,
downstream lines and devices must be free of oil and
grease.
Connect a pressure relief valve to the pressurized gas
cylinder before opening it! Before connecting the
pressure relief valve, check the connection sleeve of the
pressurized gas cylinder for contamination. Contaminated
connections must be cleaned with a clean cloth. The valve
of the pressurized gas cylinder may be opened only if the
connected fittings are closed. Open the valve of the
compressed gas cylinder very slowly.