7 Reference
7.4 Conductivity measurement
D-52/53/54/55
195
●
Conductivity and temperature coefficients for various
solutions
The following table shows the conductivity (converted to
25 ºC) and the temperature coefficient at that time, for
various kinds of solution.
Sub-
stance
Temp.
(ºC)
Conc.
(wt%)
Cond.
(S/m)
Temp.
coef.
(%/°C)
Sub-
stance
Temp.
(ºC)
Conc.
(wt%)
Cond.
(S/m)
Temp.
coef.
(%/°C)
NaOH
15
5
19.69
2.01
NaCl
18
5
6.72
2.17
10
31.24
2.17
10
12.11
2.14
15
34.63
2.49
15
16.42
2.12
20
32.70
2.99
20
19.57
2.16
30
20.22
4.50
25
21.35
2.27
40
11.64
6.48
Na
2
SO
4
18
5
4.09
2.36
KOH
15
25.2
54.03
2.09
10
6.87
2.49
29.4
54.34
2.21
15
8.86
2.56
33.6
52.21
2.36
Na
2
CO
3
18
5
4.56
2.52
42
42.12
2.83
10
7.05
2.71
NH
3
15
0.1
0.0251 2.46
15
8.36
2.94
1.6
0.0867 2.38
KCl
18
5
6.90
2.01
4.01
0.1095 2.50
10
13.59
1.88
8.03
0.1038 2.62
15
20.20
1.79
16.15
0.0632 3.01
20
26.77
1.68
HF
18
1.5
1.98
7.20
21
28.10
1.66
4.8
5.93
6.66
KBr
15
5
4.65
2.06
24.5
28.32
5.83
10
9.28
1.94
HCl
18
5
39.48
1.58
20
19.07
1.77
10
63.02
1.56
KCN
15
3.25
5.07
2.07
20
76.15
1.54
6.5
10.26
1.93
30
66.20
1.54
−
−
−
Содержание D-25
Страница 2: ......
Страница 11: ...CONTENTS VIII HORIBA ...
Страница 31: ...1 Overview of the Meter 1 9 Using the protective cap D 53 54 55 20 HORIBA ...
Страница 75: ...2 Taking Measurements 2 9 Dissolved oxygen DO measurement D 55 64 HORIBA ...
Страница 139: ...4 RS 232C communications 4 5 Communication example using the HyperTerminal 128 HORIBA ...