6 Maintenance and Troubleshooting
6.2 ION electrode maintenance
144
HORIBA
●
Before using
Before using an electrode, condition the electrode
according to the following table to prepare it for
measurement.
ION electrode
Conditioning agent
Time
Cl
-
ION electrode
No conditioning
F
-
ION electrode
NO
3
-
ION electrode
1 mol/L potassium nitrate
solution (100 g/L KNO
3
)
Approx. 1 hr
K
+
ION electrode
0.1 mol/L potassium
chloride solution (75 g/L
KCl)
Approx. 12
hr
Ca
2+
ION electrode
Tap water
Approx. 3 hr
NH
3
ammonia
electrode
No conditioning