background image

Step 1:

- Remove the mast from the box (the mast 

is 

composed of an 8000 N ac a lifting 

arm + a handlebar)

- Remove and unwrap the shin-block from 

the box

- Remove the base from the box (the base 

is equipped with anti-slip)

Step 2:

- Insert the mast in the base 

- Screw a M8 bolt in each hole of the base 

- Put the nuts covers

1

2

Assembling Instructions

BEFORE USING YOUR STAND-UP LIFT, IT IS NECESSARY TO CHECK :

• The wheels turn and roll normally.

• The rear wheels function correctly.

• There is no wear or deformation on the hooks.

WARNING :

• To preserve the inserts, do not overtighten.

• Assembly is done with the brakes locked.

User Manual - Stand-Up Lift: BLUE WAYUP PLUS with electrical opening legs

Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)

2

Содержание BLUE WAYUP PLUS

Страница 1: ...ISO 10535 2007 STAND UP LIFT BLUE WAYUP PLUS with electrical opening legs User Manual Stand Up Lift BLUE WAYUP PLUS with electrical opening legs Manufactured by NAUSICAA M dical S A S Approved by Ghiz...

Страница 2: ...Use Stand up Lifts Use Motorisation Informations Users Guide Slings Use Reuse Labelling Preventive Maintenance Security Controls General Safety Guidelines Technicals Dimensionals Characteristics Spar...

Страница 3: ...air It is permitted to be used in wet areas For example this includes the toilet or the bath This does not include using the patient hoist under the shower The stand up lift BLUE WAYUP PLUS with elect...

Страница 4: ...the base Put the nuts covers 1 2 Assembling Instructions BEFORE USING YOUR STAND UP LIFT IT IS NECESSARY TO CHECK The wheels turn and roll normally The rear wheels function correctly There is no wear...

Страница 5: ...hand wheel on the side Your stand up lift Is ready for use Assembling Instructions User Manual Stand Up Lift BLUE WAYUP PLUS with electrical opening legs Manufactured by NAUSICAA M dical S A S Approv...

Страница 6: ...bating against osteoporosis and all disorders associated with immo bility such as falls during transfers stimulate brain activity thus favouring maintenance of continence improves intestinal activity...

Страница 7: ...at can be raised by the stand up lift Do not force the controls and features of the stand up lift they are all easy to use and do not require strength Operate the stand up lift by pushing the handles...

Страница 8: ...se the castors press the parking lever pointing upwards Castor Released When locking the castors always insure that both castors are locked If only one castor is locked on a sloping floor the unlocked...

Страница 9: ...roved by Ghizlane Labrosse Biomedical Engineer Before use check that the actuator and the hand control are correctly connected to the control unit page 22 Check the status of the control box and the b...

Страница 10: ...with electrical opening legs Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer Down Closing Opening Down Overload Maintenance Battery level indication Up Up Clos...

Страница 11: ...ach sling In order to choose the right sling the caregiver shall take into account the patient s waist size All our slings provide sustainable solutions for reassuring transfers and are compatible wit...

Страница 12: ...plastron Hook the two fasteners of the backrest to the stand up lift hooks SETTINGS The fastening system with multiple hangers allows to adjust the position of the person Operate the actuator until h...

Страница 13: ...pproved by Ghizlane Labrosse Biomedical Engineer Labelling on the box Reuse The stand up lift is suitable for repeated use The stand up lift must be cleaned disinfected and serviced according to the P...

Страница 14: ...ORL Y MAINTAINED BATTERY DETERIORATES SAFETY INSTRUCTIONS BEFORE EACH USE Verify the presence and conditions of safety devices all axis and their constrained fixations pins clips and bolts and the tig...

Страница 15: ...Labelling 1 5 4 3 2 User Manual Stand Up Lift BLUE WAYUP PLUS with electrical opening legs Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 13...

Страница 16: ...l S A S Approved by Ghizlane Labrosse Biomedical Engineer A Operation of the actuator B Operation of the emergency system C Batteries status charge holding D Handcontrol status E Control operation on...

Страница 17: ...lectrical opening legs Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 1 Connection head of the actuator bolted 2 Connection base of the actuator bolted 3 Conn...

Страница 18: ...adequate measures and instructions that mechanical stress of the charging cord through bending tension shear crushing is excluded during loading or cleaning of the part This also applies to the electr...

Страница 19: ...d the SCP at its Maximal Height 880 l Minimal Distance Between the Wall and the SCP at its Maximal Height 360 m Maximal Diameter 1140 SCP Suspension central point Frame made in steel Laser cutting and...

Страница 20: ...EMB 8000N LIFTING ACTUATOR BASE SCREED all models 15 00 146 VS VRN 8 CAR 8000N all models AND 3000N VSNLV VRN LIFTING ACTUATORS COMPLETE CASING 15 00 147 VS VRN 6 8 CRD 6000N AND 8000N LIFTING ACTUATO...

Страница 21: ...cility Biological cleaning operation Steam cleaning of various flat surfaces Change surfaces regularly when washing to avoid water load Steam cleaning of inaccessible areas For tubes steam use with a...

Страница 22: ...Protection class II Protections types Remote control IP54 Control box IP65 Battery pack IP65 Actuator IP55 Stop limit switch Box in ABS Handcontrol with spiral cable low voltage 24V Protection agains...

Страница 23: ...mpt to lift a person weighing beyond the maximum weight limit is performed When it occurs the actua tor will stop and an anvil will appear it will be impossible to continue 3 Overloading of the actuat...

Страница 24: ...e battery and the control box 3 Batteries are unloaded 4 The remote control jack is not plugged 5 The remote control jack is deteriorated 6 There are damages on the remote control or on the control bo...

Страница 25: ...ection alteration or transformation not validated by NAUSICAA M dical S A S the impact of external elements natural disaster fire shock humidity flood lightning installation and or use in a non compli...

Страница 26: ...User Manual Stand Up Lift BLUE WAYUP PLUS with electrical opening legs Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 24...

Страница 27: ...User Manual Stand Up Lift BLUE WAYUP PLUS with electrical opening legs Manufactured by NAUSICAA M dical S A S Approved by Ghizlane Labrosse Biomedical Engineer 25...

Страница 28: ...66 717 184 Fax 33 0 466 532 305 Email export nausicaa medical com Website www nausicaa medical eu Manufactured by NAUSICAA M dical S A S 12 All e du Piot ZA P le Actif 30660 Gallargues le Montueux FRA...

Отзывы: