Impedance Check
An impedance check is performed to ensure that the electrode contact with the patient is satisfactory. You can perform an
impedance check at any time during a study.
Running an Impedance Check from the Software
When an impedance check is initiated, the software scans all channels (in auto scan mode).
To…
Do this…
Start the impedance check
Choose
Controls > Impedance Check
Lock onto a channel
Click
Lock Channel
. Then adjust the electrode connection until
satisfactory levels are achieved
Proceed to a full impedance check
Click
Release Lock
End the impedance check
Click
End
Save the impedance check as part of the study
Click
End and Start Recording
An impedance check displays bar graphs that show the impedance of each electrode connection. A
green
bar indicates that the
reading is below the set threshold. A
red
bar indicates the reading is above the threshold. To set the impedance threshold, click
one of the
Threshold
buttons in the
Threshold
Group
box on the right side of the
Check Impedance
box.