4.4.1
Fitting the eartip on the probe
1. Select an eartip that fits the patient's ear canal. You may have to try out a number of sizes in order to select the appro-
priate size.
2. Gently push the eartip onto the probe tip until it rests firmly against the base of the probe. It is much easier to attach
and remove the eartip if you turn it gently. When you do so, make sure that you hold the probe by the probe body
and
not by the cable
.
Note
•
Accurate testing is only guaranteed if you use the eartips provided.
Note
•
The eartip can be used for both ears. If you suspect infection in one ear, change the eartip and clean the probe
tip, or replace it with a spare, before you continue testing on the other ear.
Warning
•
Using a probe with an unsuitably sized eartip or applying excessive force may irritate the ear canal.
Note
•
In case of a probe error, make sure that the probe tip channels are clear (see
)
and that the probe is connected.
See also
.
4.4.2
Inserting the probe with eartip in the patient’s ear canal
1. When you have fitted an eartip on the probe, gently pull the pinna back and slightly
down and insert the probe in the ear canal, with a slight pressure, twisting the probe
slightly as you insert it.
Verify visually the correct fitting.
The probe can be inserted with the probe cable pointing either upwards or down-
wards, depending on which direction fits best.
Make sure that the probe fits well. Any leakage may increase the test duration
because of sound leakage, excessive noise or both.
Attach the clip to the patient's clothing or bedding to secure the probe cable.
24
Otometrics - MADSEN Alpha OAE+
4 Preparing for testing