3.2
ENG – Check Electrodes
Impedance is measured at each skin-electrode interface. A question mark (?) indicates an unacceptable electrode con-
nection.
A. Select a protocol in the
New Test
tab, then click
F7 Electrode Test
.
B. Read the status of each active electrode and channel.
Electrode Test Dialog Box
1. Check each electrode connection.
2. Check mV or K Ohm values.
3. Click
OK
.
C. Reapply electrodes or leads as needed.
D. Click
OK
to close the dialog.
E. Go to
.
16
Otometrics - ICS Chartr 200 VNG/ENG
3 Prepare Patient for Testing