Страница 2: ...Fax 1 650 802 6620 E mail customer_service natus com Technical Service 1 800 303 0306 Technical Service Fax 1 650 802 8680 E mail technical service natus com International Support Please contact your...
Страница 3: ...er Controls 11 3 1 Light Enclosure 11 3 2 Arm optional 11 3 3 Rollstand optional 12 3 4 Panel Controls 13 4 Assembly and Operating Instructions 14 4 1 Assembly 14 4 2 Setting up the device 14 4 3 Oper...
Страница 4: ...therapy for treatment guidance please refer to the AAP Guidelines American Academy of Pediatrics Clinical Practice Guideline Management of Hyperbilirubinemia in the Newborn Patient 35 or More Weeks of...
Страница 5: ...so equipped with a white exam light feature that provides approximately 10 000 lux of neutral white light for general examinations LEDs have minimal light output degradation over their lifetime with p...
Страница 6: ...he following accessories are included with each light a USB Flash Drive containing the User Manual and Service Manual and a power cord Arm optional The neoBLUE compact Arm is designed to attach to pol...
Страница 7: ...an be used alone with the neoBLUE compact Arm and with the neoBLUE compact Rollstand All configurations were tested for safety and stability and no differences were noted between various configuration...
Страница 8: ...l equipment Warning Attachment of Arm with Pole Mount When attaching the light to any floor stand other than the neoBLUE compact Rollstand the user must validate that it conforms to Regulatory Safety...
Страница 9: ...lue light off and white exam light on when checking the baby s condition and visualizing skin color blue light can hinder clinical observations by masking skin color changes such as cyanosis Monitor p...
Страница 10: ...ted is not recommended and could result in poor performance of this product Only use cables and accessories provided by Natus Medical Incorporated Warning To avoid the risk of electric shock this equi...
Страница 11: ...d put into service according to the EMC information provided in this document Important The light was factory calibrated with the neoBLUE Radiometer to deliver intensive phototherapy at a distance of...
Страница 12: ...ized European Representative Date of Manufacturing Legal Manufacturer Disposal at end of operating life instructions Catalog Number Serial Number Lot Number Atmospheric Pressure Limitation Humidity Li...
Страница 13: ...omponents and substances that may be hazardous and present a risk to human health and the environment when WEEE is not handled correctly Therefore end users also have a role to play in ensuring that W...
Страница 14: ...ases to operate contact Natus Technical Service or your authorized service provider 3 2 Arm optional The neoBLUE compact LED Phototherapy System can be mounted to the optional neoBLUE compact Arm for...
Страница 15: ...ing the gooseneck portion of the Arm and or by moving the location of the Arm clamp up or down the Rollstand Locking Casters Once the light is in place for phototherapy these casters should be locked...
Страница 16: ...after being turned off or switched to the white exam light On Standby Button Use this button to turn the blue treatment light On or to Standby Treatment Timer Reset Button The treatment timer can be...
Страница 17: ...tructions supplied with the Rollstand 4 2 Setting up the device To use the light on top of an incubator without an Arm or Rollstand follow these steps 1 Attach the light to the top of the incubator Po...
Страница 18: ...h the Arm to the manufacturer s specified location on the incubator or radiant warmer For use with the neoBLUE compact Rollstand follow the steps below 3 Attach the power cord Attach the power cord in...
Страница 19: ...your institution s procedures see Section 6 1 Checking the Light Intensity The intensity of the light was factory calibrated to deliver 35 W cm2 nm at the High setting and 15 W cm2 nm at the Low setti...
Страница 20: ...l joint near the light enclosure The light can be positioned vertically by moving it along the pole 4 Turn on the blue treatment light Press On Standby button on the control panel 5 Select High or Low...
Страница 21: ...n on white exam light as needed Press exam light button on 9 Turn off white exam light when finished Press exam light button off or press the On Standby button to resume blue treatment light 10 Turn o...
Страница 22: ...us Technical Service or authorized service provider if problem persists The light turns on but the fan is off Defective fan Defective wiring Fan is jammed due to debris Clean fan Refer to Section 6 3...
Страница 23: ...djustments is per hospital protocol Important The light was factory calibrated with the neoBLUE Radiometer to deliver intensive phototherapy at a distance of 35 cm 13 75 in from the baby The intensity...
Страница 24: ...pes Clorox Germicidal wipes Sporicidin 5 bleach Note To ensure correct operating temperature the vent filter on the underside of the light should be kept clear and free of dust As part of routine main...
Страница 25: ...central area of the effective surface area for phototherapy The intensity of the light is inversely related to the distance from the light source to the baby The light output can be adjusted to provid...
Страница 26: ...ratio 0 4 minimum to maximum within effective surface area Heat output at 35 cm 13 75 in 3 F 1 7 C warmer than ambient on mattress over 6 hrs surface White Exam Light Color Temperature Approx 4300K I...
Страница 27: ...pressure 1000 feet to 20 000 feet 50 kPa to 106 kPa Regulatory Standards IEC 60601 1 Edition 3 1 IEC 60601 2 50 Edition 2 IEC 60601 1 2 Edition 4 EMC IEC 60601 1 6 Edition 3 1 Usability Medical Applie...