Troubleshooting
Problem
ICS AirCal Overheating Warning sounds an aud-
ible continuous beep.
Causes:
1.
Water Reservoir is low/empty.
2.
ICS AirCal has been tilted 20° or more
while the pump was operating, there is a
risk of that the pump took in air instead of
water.
3.
ICS AirCal is not functioning properly.
Option 1
:
Look at the Water Level Display. If the bar is orange, this means you
need to refill the water in the ICS AirCal. Turn off the ICS AirCal. Fol-
low the Filling and Priming procedure in
. Turn on unit and verify that it is working properly.
Option 2
:
Look at the Water Level Display. If the bar is blue, then you should
have sufficient water but it should be verified. Turn off the ICS AirCal.
Follow the Filling and Priming procedure in
However, visually see if there is water present in the reservoir. If
there is no or very low water refill the water. Turn on unit and verify
that it is working properly.
Whether the units works or not, contact Natus Medical Denmark ApS
technical support and report that the ICS AirCal is not performing as
expected.
Water Removal Procedure
1.
Turn the ICS AirCal off and unplug the ICS AirCal from
the electrical wall outlet.
2.
Wipe off any dust on the top of the ICS AirCal prior to
opening the case to refill the reservoir.
3.
Face the Irrigator Front Panel. On the back right side
is a small cover. Using the Allen key (supplied with
the cradle) unscrew the cover and remove.
4.
Using the Allen key unscrew the internal cover and
remove.
5.
Using the Syringe Kit provided, remove the water
from the reservoir.
6.
Plug the ICS AirCal into the electrical wall outlet and
turn ICS AirCal on.
7.
Turn off the airflow.
8.
To prime the pump, locate the priming switch on the
back panel.
9.
Press the priming switch to turn on and hold 5
seconds.
12
Firmware procedures
To upgrade the ICS Aircal firmware, locate the ICS Aircal Firmware upgrade media and a computer on which the ICS
Aircal device driver has been installed. To install the device driver, follow the instructions in
before continuing.
Caution
• If shipping the
device, make sure to:
1) Drain the water from the
reservoir to prevent the
possibility of damage
caused by freezing, and
2) Secure the air pump
housing with the
transportation screws
supplied with the device.