Moving a Trace
In the scope window, the traces can be moved up and down for adjustment of the vertical offset.
•
Click on the
trace handle
(A) at the right end of the curve while pressing and holding down the mouse button. The trace changes
its color, and the hand pointer indicates that the trace may be moved. Finish by releasing the button at the wanted destination.
Impedance
Starting the Impedance Test
•
Click on Impedance in the right-click menu or Press
Ω
on the Amplifier box (not available for the Focus 3-channel Amplifier). The
impedance test starts on all the connected electrodes.
•
Adjust the electrodes until an acceptable level is reached (defined in the Settings window). The colored circles will change from
red to green. On the Amplifier, an acceptable electrode impedance level is indicated by all electrode input LEDs being off. An
unacceptable level (high impedance) is indicated by constantly lit LEDs.
Saving the Impedance Test
•
Click
Save
to add the impedance test to the Saved list.
Stopping the Impedance Test
•
Click
OK
to stop the impedance test.