Hardware Instructions for Use
38
Maintenance
Cleaning Instructions
The maintenance that can be performed by the operator is limited to cleaning the device. Any
maintenance inside the device must be performed by qualified service personnel only.
Regular cleaning maintenance should be performed according
to frequency of use of the
device. Always observe your local hygiene authority’s guidelines, and the following points
below:
Disconnect the mains power before cleaning the equipment.
Clean the equipment with a clean, slightly damp cloth, and wipe it dry.
Make sure that no liquids enter the device at push buttons and other openings in the
enclosure.
CAUTION
Do not use cleaning detergents, or cleaning agents based on alcohol, solvent,
silicon-based, abrasive and/or flammable substances.
Safety Checks
The following safety checks should be conducted by qualified personnel only at least once a
year and in the event of repair:
1.
Inspection for visible damage to device.
2.
Inspection of mains cord and connecting cables.
3.
Check of electrode cables and patient connections.
4.
Check of current stimulator output in all ranges –Ref. KPTest4 Manual.
5.
Measurement of insulation resistance.
6.
Measurement of leakage currents.
7.
Measurement of resistance of protective earth conductor.
8.
Measurement of resistance of protectively earthed enclosure and cart parts.
9.
Check of Amplifier gain –Ref KPTest4 Manual.