IX. Reference
ALGO
®
3i Newborn Hearing Screener IFU 024913 RevD Revision Date 08/05/2020
43
Error Messages
Error messages are most likely to occur during Start-up or Pre-screening self tests, during the
screening process, or during Exporting. Following is a table describing ALGO 3i screener Error
Messages and a Reference to the Manual and section where more information is available.
Error Message
Chapter/Section reference
Errors during Start-up or Pre-screening self tests
• ATA Calibration Data cannot be read
V. Power on/start up warnings
• ATA Calibration Due
VIII. ATA Calibration Info
• ATA Calibration Past Due
VIII. ATA Calibration Info
• ATA Calibration shutdown error
VIII. ATA Calibration Info
• Memory Error
V. Power on/start up warnings
• Battery too low shutdown error
VIII. Equipment Checks
• Battery too low to screen
VIII. Equipment Checks
• Password Incorrect
VIII. Data Protection/Passwords
Errors during screening process
• Cable Disconnected
V. Pause and Cancel Screening
• Battery charger connected
V. Pause and Cancel Screening
• Screening time limit exceeded
V. Pause and Cancel Screening
• Battery too low to complete screening
V. Pause and Cancel Screening
• Impedance Too High error
V. Prepare Baby/Check Impedances
IX. Impedance Problems
• Excessive Ambient Noise error
IX. Ambient Interference
• Excessive Myogenic Interference
IX. Myogenic Interference
• Non-physiologic Noise Interference
IX. Non-physiologic Noise
• Leads are Reversed error
IX. Leads Reversal
Errors during Exporting
• Error exporting records
Technical Notice: ALGO 3i Export Option.doc
Safety and Compliance Information
The ALGO 3i screener complies with the following safety standards:
IEC 60601-1:2005 Ed. 3+A1: C1:2014
IEC 60601-1-6:2010, A1:2013
IEC 62304:2006 Ed.1 +A1
IEC 60601-2-40:2016
IEC 60601-1-2: 2014 (EMC)
Biocompatibility: ISO 10993-1
Natus Authorized Representative information is provided on the CD which contains
this user manual.