background image

CALIBRATION PROCEDURE

NI USB-443

x

This document contains information about calibrating the National Instruments USB-4431 and 
USB-4432. These devices are referred to in this document as NI USB-443x. For more information about 
calibration, visit 

ni.com/calibration

.

Contents

Software Requirements........................................................................................................................ 2
Documentation Requirements.............................................................................................................. 2
Calibration Interval .............................................................................................................................. 2
Test Equipment .................................................................................................................................... 3
Test Conditions .................................................................................................................................... 4
Calibration Procedures......................................................................................................................... 4

Initial Setup.................................................................................................................................. 4
Changing Coupling on NI USB-443x Devices............................................................................ 5
Verification .................................................................................................................................. 6

Analog Input Performance Verification .............................................................................. 6

AI Offset Verification ................................................................................................. 6
AI AC Coupled Gain Accuracy Verification .............................................................. 7
AI DC Coupled Gain Accuracy Verification .............................................................. 8

Analog Output Performance Verification............................................................................ 10

AO Offset Verification for the NI USB-4431 Device  ................................................ 10
AO Gain Verification for the NI USB-4431 Device  .................................................. 11

Timebase Frequency Accuracy Verification ....................................................................... 12

Adjustment................................................................................................................................... 13

Analog Input Adjustment..................................................................................................... 13
Analog Output Adjustment for the NI USB-4431 Device................................................... 16

Where to Go for Support ..................................................................................................................... 18

ni.com/manuals

De

u

tsch

Français

Содержание USB-443 Series

Страница 1: ...tion Procedures 4 Initial Setup 4 Changing Coupling on NI USB 443x Devices 5 Verification 6 Analog Input Performance Verification 6 AI Offset Verification 6 AI AC Coupled Gain Accuracy Verification 7...

Страница 2: ...ed guides provides instructions for installing and configuring NI USB 443x devices NI DAQmx C Reference Help contains the C reference and general information about measurement concepts NI USB 443x Spe...

Страница 3: ...Function Generator HP Agilent 33250A Timebase Frequency Accuracy Frequency Range up to 10 kHz Frequency Accuracy 5 ppm Voltage Range up to 9 Vpk Digital Multimeter NI 4071 AO Gain AO Offset DC Voltag...

Страница 4: ...e device and configure it in Measurement Automation Explorer MAX 2 Verification Verify the existing operation of the device This step confirms whether the device is operating within the published spec...

Страница 5: ...re is no 30 seconds wait between setting the coupling on the device and starting the acquisition For flexibility NI DAQmx gives you the ability to download channel configuration settings to hardware a...

Страница 6: ...f NI USB 443x devices Refer to the NI USB 443x Specifications for the number of channels and performance specifications AI Offset Verification Note Skip the AC coupled portion of this procedure if bot...

Страница 7: ...lication 7 Repeat steps 1 through 6 In step 2 set the coupling parameter to AC Note Skip the AC coupled portion of this procedure if both of the following apply Your device is hardware revision C or e...

Страница 8: ...ifying only the channels used in your application AI DC Coupled Gain Accuracy Verification Complete the following steps to verify DC coupled gain accuracy 1 Connect the output of the calibrator to the...

Страница 9: ...1 through 5 In step 2 output 9 VDC instead of 9 VDC Subtract the averaged value measured for 0 VDC from the 9 VDC measurement Table 7 AI DC Coupled Gain Accuracy Verification Configuration for NI USB...

Страница 10: ...fy the AO offset 1 Connect channel AO 0 on the device to the input of the multimeter 2 Create and configure an AO voltage task on the NI USB 4431 as shown in Table 10 3 Create an array of length 96 00...

Страница 11: ...nfigured to generate samples continuously 5 Wait 100 ms This compensates for the filter delay of the analog output channel 6 Measure the DC voltage of the analog output channel using the multimeter in...

Страница 12: ...evice 2 Configure the function generator to generate a 10 kHz sine wave with an amplitude of 9 Vpk and no DC offset 3 Create and configure an AI voltage task on the NI USB 443x as shown in Table 14 4...

Страница 13: ...t the output of the calibrator to all input channels on the device Use the BNC T connectors to route the signal to all of the inputs 2 Configure the calibrator to output a 1 kHz signal with a 6 3 Vrms...

Страница 14: ...for NI USB 443x Configuration Value Minimum value 10 0 Maximum value 10 0 Terminal configuration Pseudodifferential Acquisition mode Continuous Rate 1000 0 Coupling AC Physical channel Dev_name aix D...

Страница 15: ...cel if there has been any error during the AI calibration or if you do not want to save the new AI calibration constants in the device EEPROM Use the action commit to save the new AI calibration const...

Страница 16: ...tinuously 6 Wait 100 ms This compensates for the filter delay of the analog output channel 7 Measure the DC voltage of channel AO 0 on the device using the multimeter Record this value as actualHighVo...

Страница 17: ...device EEPROM The analog output of the device is now calibrated with respect to your external equipment After calibrating the device verify the analog output operation by repeating the Analog Output P...

Страница 18: ...rt export data 2011 National Instruments Corporation All rights reserved 373489A 01 Oct11 Where to Go for Support The National Instruments Web site is your complete resource for technical support At n...

Отзывы: