SCXI-1141/1142/1143 Calibration Procedure
16
ni.com
3.
Call
SCXI_Set_Gain
to configure the module to a gain value of 1.
Set the following parameters:
•
SCXIchassisID
—The value obtained from MAX
•
moduleSlot
—4
•
moduleChan
—0
•
gain
—1
4.
Call
SCXI_Configure_Filter
to enable the filter bypass.
Set the following parameters:
•
SCXIchassisID
—The value obtained from MAX
•
moduleSlot
—4
•
channel
—0
•
filterMode
—0
•
freq
—0 kHz
•
cutoffDivDown
—0
•
outClkDivDown
—0
5.
Connect the calibrator to analog input channel 0. Refer to Table 7 to
determine the pins on the 96-pin front connector that correspond to the
positive and negative inputs of the specified channel. If you are using
an SCXI-1304 that is connected to the SCXI-1141/1142/1143, connect
the calibrator to the AI 0 + and AI 0 – inputs.
6.
Connect the DMM to the output of channel 0. Refer to Table 8 to
determine the pins on the 50-pin rear connector that correspond to the
positive and negative outputs for the specified channel. For example,
the positive output for channel 0 is pin 3, which is AI 0 +. The negative
input for channel 0 is pin 4, which is AI 0 –.
7.
Set the calibrator voltage to 3.4 V
rms
, 400 Hz.
8.
Measure and record the amplitude of the sine wave with the DMM.
9.
Call
SCXI_Configure_Filter
to enable the cutoff filter. Set the
following parameters:
•
SCXIchassisID
—The value obtained from MAX
•
moduleSlot
—4
•
channel
—0
•
filterMode
—1
•
freq
—25 kHz
•
cutoffDivDown
—0
•
outClkDivDown
—2