Specifications
This appendix lists the specifications for the SCXI-1120 and SCX-1120D.
These are typical at 25 °C unless otherwise stated. The operating
temperature range is 0 to 50 °C.
SCXI-1120
Analog Input
Gain (jumper-selectable)........................ 1, 2, 5, 10, 20, 50, 100, 200,
250, 500, 1,000, 2,000
Output range........................................... ±5 V
Number of channels ............................... 8
Gain accuracy......................................... 0.2% of reading typ
0.6% of reading max
Nonlinearity ........................................... 0.04% of fullscale range
Offset voltage .........................................
Stability versus ambient temperature
Offset drift RTI ...............................
Gain drift......................................... 20 ppm/°C
Input bias current .................................. ±80 pA
Input resistance
Normal ............................................ 1 G
Ω
Power off......................................... 50 k
Ω
Overload.......................................... 50 k
Ω
6
µ
V
3 mV
Gain
-------------
±
±
0.2
µ
V/
°
C
200
µ
V/
°
C
Gain
---------------------------
±
±