Chapter 2 Installation and Configuration
©
National Instruments Corporation 2-11 SCXI-1104/C User Manual
3. If you get the warning message
Failed to find
, followed by
module codes and then the message
Instead found: module with
ID 0Xxx
, return to the
section and make
sure the correct module is in the specified slot. Delete the incorrect
module as described in the
Measurement & Automation Explorer
section and then add the correct
module as described in the
4. If you get the warning message
Failed to find
, followed by a
module code and then the message
Slot x is empty
, check to see
if the configured module is installed in the specified slot. If not, install
the module by referring to the previous section,
SCXI-1104/C Module into the SCXI Chassis
. If the module is installed
in the correct slot, turn off the chassis, remove the module as specified
in Removing the SCXI-1104/C from an SCXI Chassis, and verify that
no connector pins are bent on the rear signal connector. Reinstall the
module as shown in
Installing the SCXI-1104/C Module into the SCXI
, ensuring the module is fully inserted and properly aligned in
the slot.
5. After checking the preceding items, return to the
section and retest your SCXI chassis.
If these measures do not successfully configure your SCXI system, see
Appendix B,
, for more information.
Removing the SCXI-1104/C
This section will show you how to remove the SCXI-1104/C from an SCXI
chassis.
Removing the SCXI-1104/C from an SCXI Chassis
Only a qualified person who has read and understands all the safety
information in this manual should remove an SCXI module, disconnect
signals, or remove the SCXI-1104/C from the chassis. You need the
following items to remove the SCXI-1104/C module:
• SCXI chassis or PXI combination chassis with the SCXI-1104/C
module(s) installed
• 1/4 in. flathead screwdriver