PXIe-4309 Calibration Procedure
|
© National Instruments
|
5
3.
Select
My System»Devices and Interfaces
, and then select the PXIe-4309 device to self
calibrate.
4.
Right-click the device and select
Self-Calibrate
.
Note
Once self-calibration is completed, maintain ambient temperature within
±1 °C of the self-calibration temperature.
Verification
The following procedures describe the sequence of operation and provide test points required to
verify the PXIe-4309. The verification procedures assume that adequate traceable uncertainties
are available for the calibration references.
Note
The verification procedure requires performing a self-calibration as a first
step unless self-verification is done immediately after an adjustment. Refer to the
section for instructions on performing a self calibration.
Warranted specifications have
for verification
before adjustment
and
for reverification
after adjustment
. Warranted specifications unaffected by
adjustment have test limits with no additional qualification.
Note
If any test fails verification or reverification, confirm that you have met the
before returning your device to NI. Refer to
for assistance in returning the device to NI.