Chapter 1
Setting up the SC-2345/2350 Carrier
©
National Instruments Corporation
1-19
SC-2345/2350 Carrier User Manual
SCC-PWR03
The SCC-PWR03 converts an external voltage of 7 to 42 VDC to +5 V and
±15 V and requires an external user-supplied DC power source. Attach the
voltage source to the screw terminals of J1 on the SCC‘-PWR03. Wire the
positive lead to the screw terminal labeled
7–42 V
and wire the negative
lead to the screw terminal labeled
GND
. If you are using the SC-2345
carrier, pass the leads through the strain relief on the front of the enclosure.
If you are using an SC-2345/2350 carrier with configurable connectors, use
a strain-relief panelette for the leads. You can power the SCC-PWR03 with
any appropriate 7 to 42 VDC source.
Figure 1-12.
Using the SCC-PWR03
Installing SCC Modules in the SC-2345/2350 Carrier
For instructions about how to install SCC modules in the SC-2345/2350
carrier and information about signal conditioning with SCC modules, refer
to the
SCC Quick Start Guide
.
1
SC-2345/2350 Carrier
2
7–42 VDC Source (User-Supplied)
3
DAQ Device
4
To Connector J24
5
Negative Lead to GND Screw Terminal
on SCC-PWR03 Connector J1
6
Positive Lead to +5 V Screw Terminal
on SCC-PWR03 Connector J1