Connecting the SC-2345/2350 Carrier to the DAQ Device
Caution
Refer to the
Read Me First: Safety and Radio-Frequency Interference
document
before removing equipment covers or connecting or disconnecting any signal wires.
Use a 68-pin shielded cable to connect the SC-2345/2350 carrier to the
E/M Series DAQ device. Use connector J24 on the SC-2345/2350 carrier.
The pin assignment of J24 is defined by the DAQ device MIO front
connector. Refer to Figure 1-9 for the location of connector J24.
The chassis ground terminal on the SC-2345 carrier and the
electromagnetic interference (EMI) gasket attached to the strain
relief of the SC-2345 carrier are for grounding a floating source
(1 mA maximum). Do
not
use these terminals as safety earth grounds.
Configuring the SC-2345 Carrier as a DAQ Accessory
Complete the following steps to run Measurement & Automation Explorer
(MAX) and configure the SCC system:
1.
Open MAX.
2.
Right-click
Devices and Interfaces
and select
Create New
.
3.
Under NI-DAQmx SCC Connector Block, select SC-2345. Click
Finish
.
4.
In the SCC Connector Block Configuration window, configure the
SCC system as follows.