Specifying CF00/CF40 CONFIG Settings
•
Resetting the panel settings (CF00)
•
Resetting or applying the LAN interface settings (CF40)
Complete the following steps to specify the CONFIG settings for CF00 or CF40
1.
Press CONFIG once (for CF00) or four times (for CF40).
The CONFIG key lights and the SET key blinks.
If you pressed CONFIG once, the ammeter displays the parameter number—CF00. The
voltmeter displays the corresponding setting—rSt. If you pressed CONFIG four times, the
ammeter displays the parameter number—CF40. The voltmeter displays the corresponding
setting—Lci/dEF/APPL.
2.
Press SET, which is blinking.
The CONFIG key lights, and the SET key, the parameter number displayed on the ammeter,
and the setting displayed on the voltmeter all blink.
3.
Press SET, which is blinking, again.
While the parameter number and the setting are blinking, no new settings will be applied
until you press SET.
When setting CF00, the measured value display appears.
When CF40 is set, the LAN LED lights in orange (for about 4 to 5 seconds) while the
setting is being confirmed and then turns green when it is complete. Press SET to exit the
CONFIG settings. The settings will be applied.
If the LAN LED turns red, set the settings again.
CONFIG Parameter Details
The CONFIG parameters are explained in detail below.
CF00 Resets the Panel Settings
This configuration parameter resets all of the settings listed below to their default values. The
following settings will be reset. When the settings are reset, the RMX programmable power
supply settings take on the same values as they do when an *RST command is received.
You can only use this parameter when the RMX programmable power supply is being used as a
master unit.
•
Output voltage
•
Output current
•
Overvoltage protection
•
Overcurrent protection
•
Undervoltage limit
•
Output on/off at power-on
•
Constant current control using an Vext or Rext (CF05)