RMX-4000 Series Programming Manual
176
OT
When the internal temperature exceeds
85
˚C
, the OT bit is set. The OT bit is
automatically cleared after the
temperature drops below 85
˚C.
G/N
The Go/NoGo bit is set when Go/NoGo
limits have been exceeded, when
Go/NoGo SPEC has been enabled.
UVP
If the under voltage condition occurs, the
UVP bit is set.
Condition
Register
The condition register indicates the electronic load
status. Only a change in the electronic load
condition can change the condition register.
Reading the condition register does not change the
condition register state.
PTR/NTR Register
The PTR/NTR (Positive/Negative transition)
register determines the type of transition
conditions that will trigger an event. You can
transition program only the Channel Status
Register and Questionable Status Register.
Positive Transition
0
1
Negative Transition
1
0
Event Register
The Event Register indicates whether an event has
been triggered according to the transition settings
from the PTR/NTR Register.
Enable Register
The Enable register determines which status
event(s) are enabled. Any enabled status events
(OC, OV, OP, RV, OT, G/N, UVP) will set the
corresponding channel bit in the Channel
Summary Event Register.
Channel Summary
Description
The Channel Summary Registers consolidate the
channel status of all 4/8 channels, depending on
the electronic load.
Содержание RMX-400 Series
Страница 1: ...Programmable DC Electronic Load RMX 400x Series PROGRAMMING MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 23: ...RMX 4000 Series Programming Manual 23 Memory Subsystem 153 SEQuence Subsystem 159 GLOBal Subsystem 170...
Страница 31: ...RMX 4000 Series Programming Manual 31 0 Pass 1 Fail Example TST 0...
Страница 37: ...RMX 4000 Series Programming Manual 37 Query Example MEM this is a memo Returns the memo message...
Страница 56: ...RMX 4000 Series Programming Manual 56 Query Example UTIL FRAM 0 Frame Link is on...
Страница 75: ...RMX 4000 Series Programming Manual 75 Syntax FETCh ALLPower aard...
Страница 96: ...RMX 4000 Series Programming Manual 96 1 ON Turn on Example OCP RUN OFF Turn the test off...
Страница 134: ...RMX 4000 Series Programming Manual 134 Questionable Status NTR All bits set to 0 Syntax STATus PREset Example STAT PRE...