background image

RM-4339 Calibration Procedure

|

© National Instruments

|

7

7.

Repeat steps through 6

 

for each channel you want to verify.

8.

Disconnect the DMM from the module.

Offset Voltage Verification

The offset voltage verification uses the NI PXIe-4339 to measure the offset voltage of the 
RM-4339. Prior to measuring the offset voltage of the RM-4339, an offset characterization is 
performed on the NI PXIe-4339 using a TB-4339/B/C to provide an electrical short on the input 
of the NI PXIe-4339. The temperature variation of the environment should be kept within 1 °C 
between measurement of the offset error of the NI PXIe-4339 and the offset measurement of the 
RM-4339.

Complete the following steps to verify the offset voltage:
1.

Make the following connections:
a.

On the TB-4339/B/C, short the AI+ and AI- of all the channels to AIGND.

Note

Refer to the 

NI PXIe-4339 and TB-4339/B/C Installation Guide and Terminal 

Block Specifications

 for the pin assignments and signal names.

b.

Connect the female end of the SH96-96-2 cable to the NI PXIe-4339.

c.

Connect the TB-4339/B/C to the male end of the SH96-96-2 cable.

Table 3.  

SCA Shunt Calibration Resistance Accuracy Limits

Nominal Value (

)

Lower Limit (

)

Upper Limit (

)

100,000

99,900

100,100

Содержание RM-4339

Страница 1: ...curacy Verification 9 Specifications 12 World Wide Support and Services 13 Software Calibrating the NI RM 4339 requires the installation of NI DAQmx on the calibration system Driver support for the RM...

Страница 2: ...dware setup NIPXIe 4339 User Manual NI PXIe 4339 specific information NI PXIe 4339 Specifications NI PXIe 4339 specifications and calibration interval NI DAQmx Readme Operating system and application...

Страница 3: ...r and can sink 15 mA on the 120 setting 8 mA on the 350 setting and 3 mA on the 1 k setting Ohm Meter NI PXI 4070 If this instrument is unavailable use a 6 digit DMM with 2 wire resistance measurement...

Страница 4: ...how to install and use the RM 4339 Test Conditions The following setup and environmental conditions are required to ensure the RM 4339 meets published specifications Keep connections to the RM 4339 a...

Страница 5: ...erification procedures assume that adequate traceable uncertainties are available for the calibration references SCA Shunt Calibration Resistance Accuracy Verification This section provides instructio...

Страница 6: ...hunt Cal c Commit the task by calling DAQmx Control Task with the following parameter Action Commit 4 Take a resistance measurement from the DMM and record the result 5 Compare the result from step 4...

Страница 7: ...n of the environment should be kept within 1 C between measurement of the offset error of the NI PXIe 4339 and the offset measurement of the RM 4339 Complete the following steps to verify the offset v...

Страница 8: ...nector on the back of the RM 4339 4 Connect an RJ50 cable to each channel with AI pin 2 and AI pin 3 connected to T pin 9 5 Acquire voltage measurements from all channels with the NI PXIe 4339 a Creat...

Страница 9: ...connect QTR pin 5 to LO and connect AI pin 2 to LO as shown in Figure 2 Figure 2 RJ50 to Calibrator Connections b QTR and AI must be connected together at the calibrator The EX lead wire and the QTR l...

Страница 10: ...erage the readings that you acquired and record the values as ResultSCD g Clear the task by calling DAQmx Clear Task 7 Acquire an AI Bridge V V measurement with shunt calibration enabled a Create a ta...

Страница 11: ...47 for 350 SCalExp 0 00248756 for 1 k For reference the calculation for the expected shunt calibration deflection is given by the following equation where Req is the equivalent parallel resistance of...

Страница 12: ...e Test limits in this document are based upon the March 2015 edition of the NI PXIe 4339 and RM 4339 User Guide and Terminal Block Specifications Refer to the most recent NI PXIe 4339 and RM 4339 User...

Страница 13: ...ivate expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014 and DFAR 252 227 7015 World Wide Support and Services The Natio...

Отзывы: