RM-26999 Calibration Procedure
12
ni.com/calibration
AC Gain Verification
Test Limits
Table 2:
AC Gain Verification Test Limits
Frequency
High
Voltage
Input Value
Low Voltage
Input Value
Calibrator
Range
DMM
Range
As-Found Test Limit
As-Left Test Limit
Lower Limit
Upper Limit
Lower Limit
Upper Limit
500 Hz
1000 V
RMS
5 V
RMS
1100 V
10 V
-1.0 V
RMS
+1.0 V
RMS
-500 mV
RMS
+500 mV
RMS
500 V
RMS
2.5 V
RMS
1100 V
10 V
-513 mV
RMS
+513 mV
RMS
-260 mV
RMS
+260 mV
RMS
50 V
RMS
0.25 V
RMS
22 V
1 V
-63 mV
RMS
+63 mV
RMS
-30 mV
RMS
+30 mV
RMS
5 V
RMS
0.025 V
RMS
22 V
0.1 V
-18 mV
RMS
+18 mV
RMS
-9 mV
RMS
+9 mV
RMS
1 kHz
1000 V
RMS
5 V
RMS
1100 V
10 V
-2.0 V
RMS
+2.0 V
RMS
-1.0 VRMS
+1.0 V
RMS
500 V
RMS
2.5 V
RMS
1100 V
10 V
-1.0 V
RMS
+1.0 V
RMS
-500 mV
RMS
+500 mV
RMS
50 V
RMS
0.25 V
RMS
22 V
1 V
-113 mV
RMS
+113 mV
RMS
-55 mV
RMS
+55 mV
RMS
5 V
RMS
0.025 V
RMS
22 V
0.1 V
- 80.5 mV
RMS
+80.5 mV
RMS
-40 mV
RMS
+40 mV
RMS
5 kHz
1000 V
RMS
5 V
RMS
1100 V
10 V
-9.0 V
RMS
+9.0 V
RMS
-4.5 V
RMS
+4.5 V
RMS
500 V
RMS
2.5 V
RMS
1100 V
10 V
-4.5 V
RMS
+4.5 V
RMS
-2.1 V
RMS
+2.1 V
RMS
50 V
RMS
0.25 V
RMS
22 V
1 V
-463 mV
RMS
+463 mV
RMS
-230 mV
RMS
+230 mV
RMS
5 V
RMS
0.025 V
RMSRMS
22 V
0.1 V
- 80.5mV
RMS
+80.5 mV
RMS
-40 mV
RMS
+40 mV
RMS
10 kHz
1000 V
RMS
5 V
RMS
1100 V
10 V
-11.5 V
RMS
+11.5 V
RMS
-5.5 V
RMS
+5.5 VRMS
500 VRMS
2.5 V
RMS
1100 V
10 V
-5.8 VRMS
+5.8 V
RMS
-2.6 V
RMS
+2.6 V
RMS
50 V
RMS
0.25 V
RMS
22 V
1 V
-588 mV
RMS
+588 mV
RMS
-290 mV
RMS
+290 mV
RMS
5 V
RMS
0.025 V
RMS
22 V
0.1 V
- 80.5mV
RMS
+80.5 mV
RMS
-40 mV
RMS
+40 mV
RMS
15 kHz
1000 V
RMS
5 V
RMS
1100 V
10 V
-13.5 V
RMS
+13.5 V
RMS
-6.5 V
RMS
+6.5 V
RMS
500 V
RMS
2.5 V
RMS
1100 V
10 V
-6.8 V
RMS
+6.8 V
RMS
-3.4 V
RMS
+3.4 V
RMS
50 V
RMS
0.25 V
RMS
22 V
1 V
-688 mV
RMS
+688 mV
RMS
-340 mV
RMS
+340 mV
RMS
5 V
RMS
0.025 V
RMS
22 V
0.1 V
- 80.5mV
RMS
+80.5 mV
RMS
-40 mV
RMS
+40 mV
RMS