Voltage (V)
Current (Amps) Typical
Current (Amps) Maximum
-12 V
0 A
0 A
+5 V
Aux
0.19 A
0.46 A
Physical
Board dimensions
Four-wide 3U PXI Express module
Slot requirements
One system slot plus three controller expansion
slots
Compatibility
Fully compatible with
PXI Express
Specification
1.0
Weight
0.98 kg (2.16 lb) typical
Environmental
Maximum altitude
2,000 m (800 mbar) (at 25 °C ambient)
Pollution Degree
2
Indoor use only.
Operating Environment
Caution
The operating temperature must not be exceeded, even when used in a
chassis with a higher temperature range.
Ambient temperature range
5 °C to 50 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 high temperature
limit.)
Extended temperature option
0 °C to 55 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 low temperature
limit and MIL-PRF-28800F Class 2 high
temperature limit.)
Relative humidity range
10% to 90%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
22
|
ni.com
|
PXIe-8821 User Manual