Documentation for the NI-IMAQdx driver software is available by selecting
Start
»
All
Programs
»
National Instruments
»
Vision
»
Documentation
»
NI-IMAQdx
.
3.
NI Vision Development Module—Refer to the
NI Vision Development Module Readme
on the NI Vision Development Module installation media for system requirements and
installation instructions.
Documentation for Vision Development Module is available by selecting
Start
»
All
Programs
»
National Instruments
»
Vision
»
Documentation
»
NI Vision
.
Installing the PXIe-8246
Notice
To prevent damage to the PXIe-8246 caused by ESD or contamination,
handle the module using the edges or the metal bracket.
1.
Ensure the AC power source is connected to the chassis before installing the module.
The AC power cord grounds the chassis and protects it from electrical damage while you
install the module.
2.
Power off the chassis.
3.
Inspect the slot pins on the chassis backplane for any bends or damage prior to
installation. Do not install a module if the backplane is damaged.
4.
Remove the black plastic covers from all the captive screws on the module front panel.
5.
Identify a supported slot in the chassis. The following figure shows the symbols that
indicate the slot types.
Figure 1. Chassis Compatibility Symbols
NI PXIe-1062Q
1
2
3
4
5
1. PXI Express System Controller Slot
2. PXI Peripheral Slot
3. PXI Express Hybrid Peripheral Slot
4. PXI Express System Timing Slot
5. PXI Express Peripheral Slot
6.
Touch any metal part of the chassis to discharge static electricity.
7.
Place the module edges into the module guides at the top and bottom of the chassis. Slide
the module into the slot until it is fully inserted.
8.
Secure the module front panel to the chassis using the front-panel mounting screws.
Note
Tightening the top and bottom mounting screws increases mechanical
stability and also electrically connects the front panel to the chassis, which can
improve the signal quality and electromagnetic performance.
9.
Cover all empty slots using EMC filler panels or fill using slot blockers to maximize
cooling air flow, depending on your application.
10. Connect an Ethernet cable to a GigE Vision camera, then connect the camera to the