2.
Power off the PXI Express chassis.
3.
Remove the module from the PXI Express chassis and inspect it for damage. Do not
reinstall a damaged device.
4.
Install the module in a different PXI Express chassis slot from which you removed it.
5.
Power on the PXI Express chassis.
6.
Restart your computer.
7.
Verify that the module appears in MAX.
What Should I Do if the PXIe-5830 Fails the Self-Test?
1.
Restart the system.
2.
Launch MAX, and perform the self-test again.
3.
Power off the chassis.
4.
Reinstall the failed module in a different slot.
5.
Power on the chassis.
6.
Perform the self-test again.
What Should I Do if the Instrument Does Not Initialize?
Failure to initialize may indicate a problem with module interconnections or with MAX. If the
niRFSA Initialize VI or niRFSG Initialize VI returns an error and the PXIe-5830 fails to
initialize, complete the following steps:
1.
Reconnect the PXIe-5830 front panel cables securely.
2.
Power on your system and run the MAX configuration and self-test procedures.
18
|
ni.com
|
PXIe-5830 Getting Started Guide