background image

1.

Launch MAX by navigating to 

Start

»

All Programs

»

National Instruments

»

NI MAX

 or

by clicking the NI MAX desktop icon.

2.

In the Configuration pane, expand 

Devices and Interfaces

 to see the list of installed

devices. Installed devices appear under the name of their associated chassis.

3.

Expand your 

Chassis

 tree item.

MAX lists all devices installed in the chassis. PXIe-5820 devices appear as NI-RIO
devices in the list. Your default device names may vary.

Note

 

If you do not see your hardware listed, refer to the 

Troubleshooting

section of this document.

4.

Record the device identifier MAX assigns to the hardware. Use this identifier when
programming the PXIe-5820.

Caution

 

When you install, uninstall, or move an NI-RIO device in your

system, resource identification of your NI-RIO devices may change. Whenever
any of these changes occur, verify resource identification of all your NI-RIO
devices in MAX, and, if necessary, make changes to your software and
documentation.

Self-Calibration

Self-calibration adjusts the PXIe-5820 for variations in the module environment using an
onboard high-precision calibration tone.

The PXIe-5820 modules are externally calibrated at the factory; however, you should perform
a self-calibration in any of the following situations:

After first installing the PXIe-5820 into your chassis

After any module in the chassis is installed, uninstalled, or moved

When the system is in an environment where the ambient temperature varies or the
module temperature has drifted more than ±5 °C from the temperature at the last self-
calibration

To periodically adjust for small performance drifts that occur with product aging

Note

 

Allow the module to warm up for 30 minutes after the PXI Express chassis

has powered on and the operating system has completely loaded before you use the
module or perform self-calibration.

Note

 

Disconnect external I/Q port cables prior to self calibration for best

performance. If this is not possible then make sure external devices are turned off or
in idle state and not transmitting. It is not recommended to perform self calibration
while connected in direct loopback configuration.

Performing a Device Self-Calibration Using NI-RFSA

NI recommends you perform the self-calibration using the NI-RFSA example when you self-
calibrate the module.
1.

Select 

Start

»

All Programs

»

National Instruments

»

LabVIEW

 to launch LabVIEW.

PXIe-5820 Getting Started Guide

  | 

© National Instruments

  | 

11

Содержание PXIe-5820

Страница 1: ...nes 2 Verifying the System Requirements 2 Unpacking the Kit 2 Preparing the Environment 3 Verifying the Kit Contents 3 Other Equipment 4 Choosing and Installing the Software 5 Software Options 5 Installing the Software 5 Installing the PXIe 5820 6 Direct Connections to the PXIe 5820 7 Hardware Front Panel Connectors and Indicators 8 Configuring the Hardware in MAX 10 Self Calibration 11 Performing...

Страница 2: ...aution To ensure the specified EMC performance operate this product only with high quality shielded cables and accessories Caution To ensure the specified EMC performance operate this product only with cables less than 1 meter in length Verifying the System Requirements To use the PXIe 5820 your system must meet certain requirements For more information about minimum system requirements recommende...

Страница 3: ...om contaminants before returning it to service Note Refer to the PXIe 5820 Specifications at ni com manuals for complete specifications Verifying the Kit Contents Figure 1 PXIe 5820 Kit Contents 1 2 3 5 6 4 SD NSIT IVE NI PX Ie 5 820 Vec tor Sig nal Tran sce iver Bas eba nd IQ 1 25 GS s PFI 0 DIO I I Q Q I I Q Q REF OU T REF IN ACCE SS ACTIV E I Q IN 100 Ω DIFF I Q OU T 100 Ω DIFF 50 Ω 50 Ω 1 PXIe...

Страница 4: ...ess device For more information about compatible chassis options visit ni com info and enter the Info Code pxiechassis A PXI Express embedded controller or PC with MXI controller system that meets the system requirements specified in this guide and chassis documentation PXI Chassis Slot Blocker kit NI part number 199198 01 Optional Items SMA m to MMPX m cables MMPX m to MMPX m cables OCuLink x4 1 ...

Страница 5: ...Toolkit to develop software defined radio transmitters and receivers Use NI RFSG with the NI Modulation Toolkit to create and generate modulated signals to test receivers Related Information Refer to the NI RF Vector Signal Analyzers Help or the NI RF Signal Generators Help for more information about using the NI RFSA and NI RFSG instrument driver FPGA extensions Installing the Software You must b...

Страница 6: ...om electrical damage while you install the module 2 Power off the chassis 3 Inspect the slot pins on the chassis backplane for any bends or damage prior to installation Do not install a module if the backplane is damaged 4 If the chassis has multiple fan speed settings ensure the fans are set to the highest setting Note Inadequate air circulation could cause the temperature inside the chassis to r...

Страница 7: ...up on the ejector handle 12 Secure the module front panel to the chassis using the front panel mounting screws Note Tightening the top and bottom mounting screws increases mechanical stability and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 13 Use slot blockers and cover all empty slots with filler panels to maximiz...

Страница 8: ...atic free environment If you are using an active device such as a preamplifier or switch routed to the PXIe 5820 I Q IN connectors ensure that the device cannot generate signal transients greater than the AC and DC specifications of the PXIe 5820 I Q IN connectors Hardware Front Panel Connectors and Indicators Figure 4 PXIe 5820 Front Panel NI PXIe 5820 Vector Signal Transceiver Baseband IQ 1 25 G...

Страница 9: ...external 10 MHz Reference Clock MMPX f REF OUT Output connector that can export a 10 MHz Reference Clock MMPX f PFI 0 Programmable function digital I O DIO connector for use with triggers or events MMPX f DIO Multi signal DIO connector that provides access to FPGA multi gigabit transceivers MGTs and general purpose LVCMOS signals Nano Pitch Caution The DIO connector is not an OCuLink interface Do ...

Страница 10: ...he device Green The device is controllable through the software ACTIVE Off The device is idle Solid green The device is generating a waveform Dim amber The device is waiting for an acquisition Reference Trigger Solid amber The device is acquiring a waveform Solid red The device has detected an error The LED remains red until the error condition is removed Note The indicators are listed in increasi...

Страница 11: ...The PXIe 5820 modules are externally calibrated at the factory however you should perform a self calibration in any of the following situations After first installing the PXIe 5820 into your chassis After any module in the chassis is installed uninstalled or moved When the system is in an environment where the ambient temperature varies or the module temperature has drifted more than 5 C from the ...

Страница 12: ...e a Select the device identifier assigned to the PXIe 5820 in MAX in the Resource Name drop down menu b Set Clock Source to OnboardClock c Set Self Calibration Step Operations to Perform All Self Calibration Steps 4 Run the VI Related Information Refer to the NI RF Vector Signal Analyzers Help for more information about self calibration and self calibrating in text based languages 12 ni com PXIe 5...

Страница 13: ... bit 7 64 bit Program Files x86 IVI Foundation IVI Microsoft Visual C C Use the header files located in the IVIROOTDIR32 Include directory and import library files located in the IVIROOTDIR32 Lib directory where IVIROOTDIR32 is one of the following directories Windows 10 32 bit 8 1 32 bit 7 32 bit Program Files IVI Foundation IVI Windows 10 64 bit 8 1 64 bit 7 64 bit Program Files x86 IVI Foundati...

Страница 14: ...OOTDIR32 is one of the following directories Windows 10 32 bit 8 1 32 bit 7 32 bit Program Files IVI Foundation IVI Windows 10 64 bit 8 1 64 bit 7 64 bit Program Files x86 IVI Foundation IVI Microsoft NET To use the NET API you must install the NET class libraries For download and installed file locations visit ni com info and enter Info Code NETAPIdriversupport Programming Examples Locations Usin...

Страница 15: ... Q port to the I Q OUT Q port using an MMPX m to MMPX m cable d Connect the I Q IN Q port to the I Q OUT Q port using an MMPX m to MMPX m cable 2 Select Start All Programs National Instruments LabVIEW to launch LabVIEW 3 Complete the following steps to open the Getting Started Examples a Select LabVIEW Help Find Examples b Select Hardware Input and Output Modular Instruments NI RFSA Getting Starte...

Страница 16: ...and NI RIO FPGA Device nodes and select Update Driver If you are using an MXI controller right click PCI to PCI Bridge and select Properties from the shortcut menu to verify that the bridge is enabled 6 Restart your computer Why Is the ACCESS LED Off When the Chassis Is On The LEDs may not light until the device has been configured in MAX Before proceeding verify that the PXIe 5820 appears in MAX ...

Страница 17: ...y ni com community Support ni com support more about your products through ni com Located online at ni com manuals Located online at ni com gettingstarted RF Solutions ni com rf PXIe 5820 Specifications NI RF Vector Signal Transceivers Help Tip The NI RF Vector Signal Transceivers Help available at ni com manuals is an HTML version of a traditional user manual that includes detailed information ab...

Страница 18: ... numbers email addresses and current events Information is subject to change without notice Refer to the NI Trademarks and Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering NI products technology refer to the appropriate location Help Patents in your...

Отзывы: