1.
Launch MAX by navigating to
Start
»
All Programs
»
National Instruments
»
NI MAX
or
by clicking the NI MAX desktop icon.
2.
In the Configuration pane, expand
Devices and Interfaces
to see the list of installed
devices. Installed devices appear under the name of their associated chassis.
3.
Expand your
Chassis
tree item.
MAX lists all devices installed in the chassis. PXIe-5820 devices appear as NI-RIO
devices in the list. Your default device names may vary.
Note
If you do not see your hardware listed, refer to the
Troubleshooting
section of this document.
4.
Record the device identifier MAX assigns to the hardware. Use this identifier when
programming the PXIe-5820.
Caution
When you install, uninstall, or move an NI-RIO device in your
system, resource identification of your NI-RIO devices may change. Whenever
any of these changes occur, verify resource identification of all your NI-RIO
devices in MAX, and, if necessary, make changes to your software and
documentation.
Self-Calibration
Self-calibration adjusts the PXIe-5820 for variations in the module environment using an
onboard high-precision calibration tone.
The PXIe-5820 modules are externally calibrated at the factory; however, you should perform
a self-calibration in any of the following situations:
•
After first installing the PXIe-5820 into your chassis
•
After any module in the chassis is installed, uninstalled, or moved
•
When the system is in an environment where the ambient temperature varies or the
module temperature has drifted more than ±5 °C from the temperature at the last self-
calibration
•
To periodically adjust for small performance drifts that occur with product aging
Note
Allow the module to warm up for 30 minutes after the PXI Express chassis
has powered on and the operating system has completely loaded before you use the
module or perform self-calibration.
Note
Disconnect external I/Q port cables prior to self calibration for best
performance. If this is not possible then make sure external devices are turned off or
in idle state and not transmitting. It is not recommended to perform self calibration
while connected in direct loopback configuration.
Performing a Device Self-Calibration Using NI-RFSA
NI recommends you perform the self-calibration using the NI-RFSA example when you self-
calibrate the module.
1.
Select
Start
»
All Programs
»
National Instruments
»
LabVIEW
to launch LabVIEW.
PXIe-5820 Getting Started Guide
|
© National Instruments
|
11